Volta Medical
Volta Medical makes Volta AF-Xplorer II, AI software that helps electrophysiologists decide where to ablate in patients with persistent atrial fibrillation and atrial tachycardia. Running on a dedicated computer in the EP lab, it reads intracardiac electrograms from the mapping catheter in real time and flags those showing spatiotemporal dispersion, a pattern the founders linked to the drivers that sustain AF. Electrodes over dispersed tissue turn red with an audible cue, a gauge marks highly stable dispersion, and the operator tags regions on the 3D map by hand or, with Abbott's EnSite X, automatically. Booster mode raises sensitivity during atrial tachycardia or slow AF. The classifiers are machine and deep learning models trained on 832,250 annotated electrogram samples from Europe and the United States, three times the earlier version's database, curated through VoltaPlex, Volta's data collection program with partner centers.
AF-Xplorer II received FDA 510(k) clearance in September 2025 and launched in the US that November, following VX1, cleared in 2020, and AF-Xplorer. It works with Boston Scientific LabSystem Pro and GE CardioLab recording systems, embeds in GE's CardioLab AltiX AI.i screen, and supports CARTO 3, EnSite X, Affera and Opal HDx workflows, including pulsed field catheters. The TAILORED-AF trial, a 370 patient, 26 center randomized controlled trial of the earlier Volta system published in Nature Medicine in 2025, found 88% freedom from AF at 12 months after one AI guided procedure against 70% with pulmonary vein isolation alone, with no significant difference in freedom from any atrial arrhythmia. Volta was founded in Marseille in 2016 by data scientist Théophile Mohr Durdez and electrophysiologists Julien Seitz, Jérôme Kalifa and Clément Bars, has raised more than €90 million and has staff in France, the US and Germany.
Capability Axes
An AI Health Index grade measures what a buyer can verify from public sources on the date shown. It is not a rating of how good the product is. A vendor can build an excellent system and grade low on an axis because it publishes nothing an outsider can check. How grades read
Volta AF-Xplorer II exists to run one learned function: machine learning and deep learning classifiers decide, for each bipole on the mapping catheter, whether its electrograms show spatiotemporal dispersion, and mark highly stable dispersion as a subset. The 510(k) summary for K251518 calls the device a machine and deep learning based algorithm, and the computer, cables and display are there to feed it signals and show its output.
Cycle length estimation runs beside the classifiers, but without them there is no dispersion map to tag and nothing for an electrophysiologist to buy. The models replace the visual electrogram reading that the founders' original method asked of experienced operators.
Electrophysiologists stay in charge of every target. The user manual sets out the sequence: the operator moves the catheter, stabilizes it when bipoles blink, waits for the software to confirm dispersion on the outer frame, and can hold five to 15 seconds for the highly stable check before tagging.
Tags go onto the 3D map manually with LabSystem Pro, CardioLab or any navigation system, and automatically only with Abbott EnSite X. Use is restricted to electrophysiologists trained by Volta or working under a proctor, with training after each installation and software upgrade, and booster mode for slow AF or atrial tachycardia is the operator's choice. What happens after a wrong or missed flag, and how automatic tags are reviewed or removed, rest with clinical judgment.
The dispersion classifiers are machine learning and deep learning models, trained on 832,250 electrogram samples from Europe and the United States, three times the database of the previous version, with annotations from electrophysiologists in both regions. Each version has its own clearance and user manual: VX1 in 2020, VX1+ in 2023, AF-Xplorer in 2023 and 2025, and AF-Xplorer II as K251518 in September 2025, with manuals for software versions 3.1 and 3.2 posted with their dates.
The model architectures, the input features beyond the 1.5 second electrogram window, and how VoltaPlex data feeds each retraining are never described, so the training set is characterized by size and origin rather than by patient mix.
Models are built in house, and the software is installed on a computing platform at Volta's manufacturing site and configured on site, so analysis runs on the EP lab computer rather than through a cloud model provider. The cleared device cannot change without a new clearance, and the user manual says it connects only to identified third party systems: the Boston Scientific and GE recording systems and Abbott's EnSite X. Training data comes from VoltaPlex, which collects procedural data from partner centers and has it curated by electrophysiologists, with every research record pseudonymized. Terms governing that collection, such as which centers contribute and how long data is kept, are not offered.
TAILORED-AF, a randomized controlled trial in Nature Medicine (2025), assigned 370 patients with persistent or long standing persistent AF, treated by 51 operators at 26 centers in Europe and the US, to ablation of AI identified dispersion areas plus pulmonary vein isolation or to isolation alone.
Freedom from AF at 12 months after one procedure, the primary endpoint, was 88% against 70%; freedom from any atrial arrhythmia after one procedure did not differ significantly, and ECGs went to a blinded independent core lab. The trial ran from February 2021 to December 2022 on the earlier Volta AF-Xplorer system.
AF-Xplorer II runs updated classifiers trained on a larger electrogram database, and its K251518 clearance rests on bench testing against expert assessment rather than a clinical study. RESTART (Heart Rhythm, 2026), a single arm trial in redo patients at 20 centers enrolled from March 2023 to April 2024, and Ev-AIFib (Journal of Cardiovascular Electrophysiology, 2022) also tested earlier versions, and Volta staff coauthored all three.
Procedural data from partner centers feeds VoltaPlex, where electrophysiologists curate it for training, and all data collected for research is pseudonymized; the training database is continuously enriched with new procedures to improve the algorithms. On the device, signals are analyzed locally, and the user manual says stored information is not retained, apart from logs, when a system is withdrawn and factory reset.
No retention period applies to the VoltaPlex research database, and no statement covers whether data from ordinary commercial use is collected or used for training. Safety engineering for the classifiers, such as how a misclassification is detected and reported, goes no further than the device's general warnings.
The US privacy statement covers visitors to the website and personal information Volta gathers in business, such as customer contacts, billing details and adverse event reports from physicians, and allows transfer to affiliates and vendors abroad. It makes no mention of HIPAA, business associate status or the patient signals the device processes.
The device analyzes electrograms on a local computer in the EP lab, and VoltaPlex research data is pseudonymized before Volta uses it, a commitment no privacy document for the product repeats. No business associate agreement, and no statement that Volta handles no protected health information, is offered to US hospitals.
Cybersecurity guidance sits in the user manual: the system connects only to identified third party systems, other endpoints are strongly discouraged, users must keep credentials private, only identified accessories and maintenance USB storage may be plugged in, and any physical tampering is reported to Volta. The 510(k) summary lists a Trusted Platform Module on the computing platform, and a withdrawn system is factory reset.
No SOC 2 report, ISO 27001 certificate, penetration test, software bill of materials or trust portal is offered, and no cybersecurity review appears in the 510(k) summary. The device talks to EnSite X and CardioLab over a bidirectional ethernet link.
Volta AF-Xplorer II is cleared under K251518 (September 12, 2025, product code DQK, 21 CFR 870.1425) to assist operators in manual or automatic annotation of 3D atrial maps for multipolar electrograms showing spatiotemporal dispersion during AF or atrial tachycardia.
The product page says the device is FDA cleared and quotes that indication, with no clearance number or date, and sells it as bringing expert electrophysiologists' acute and long term outcomes into the lab by detecting electrograms indicative of arrhythmia drivers. The clearance covers annotation, while the outcome and driver language rests on the trials and the founders' 2017 JACC work. Earlier versions were cleared as VX1 in 2020 and AF-Xplorer in 2023 and 2025, and the EU versions hold CE marks.
A TAILORED-AF analysis of women in Europace (2025) compared results by sex: after AI guided ablation, freedom from AF was 76% in women and 91% in men (P = 0.07), and freedom from any atrial arrhythmia 56% and 62%, while with isolation alone women did significantly worse than men. A Heart Rhythm O2 subanalysis (December 2025) reports improvement over isolation alone across NYHA heart failure classes, and the main trial reported patients with AF lasting six months or more separately.
All three come from the trial of the earlier Volta system, before AF-Xplorer II's classifiers were retrained. The training database draws on annotations from European and US electrophysiologists, and VoltaPlex names representativeness of training data as a bias risk. No accuracy for the classifier itself is reported by sex, rhythm or catheter type.
Every flag reaches an electrophysiologist before anything is ablated: the software colors bipoles and sounds a cue, the operator confirms and tags, and the ablation decision stays with the physician under the manual's requirement of clinical judgment and Volta training. Labeling states there are no known adverse events associated with the software.
No error rate for the dispersion classifier is given, since the 510(k) summary reports unit tests at 100% conformity and an expert qualitative comparison of maps rather than sensitivity or false flags. No warranty, indemnity or remedy covers a missed or wrong flag.
The integrations are with the EP lab's own systems rather than an EHR. The 510(k) and user manual name analog links to Boston Scientific LabSystem Pro and GE CardioLab, and bidirectional digital links to GE CardioLab and Abbott EnSite X through Live Export or LiveSync, where tags land on the 3D map automatically.
GE's CardioLab AltiX AI.i can embed the Volta interface side by side on one screen, and manual tagging works with CARTO 3, Affera, including Sphere-9 one catheter workflows, and Opal HDx. The connections run in procedures in the US and Europe, and no link to the hospital EHR is offered.
Volta AF-Xplorer II is delivered as a dedicated computer, installed with the software at Volta's manufacturing site in Marseille and configured on site, sitting outside the sterile field and connected only to the lab's recording and mapping systems. Electrogram analysis happens on that machine, so procedure data rests in the hospital, and a withdrawn system is factory reset with only logs kept.
VoltaPlex separately gathers pseudonymized procedural data from partner centers for training, without saying where that database is hosted or whether US data leaves the country, and the website privacy statement allows transfers to affiliates and vendors abroad.
The product is a capital system: a computing platform with software, cables, monitor and accessories, installed and configured by a Volta representative for the lab's recording or mapping system, with training for electrophysiologists after each installation and software upgrade. Software releases arrive as numbered versions with their own manuals, 3.1 in October 2025 and 3.2 in November 2025, and the list of compatible recording and mapping systems keeps growing.
No price, license term, per procedure fee or service contract figure is given, and buyers reach Volta through a contact form or a local representative. Investor material puts total capital raised above €90 million and dates the first commercial case to October 2020.
Catheter ablation of persistent and long standing persistent AF in the EP lab is the core setting, with atrial tachycardia in the indication and booster mode for slow AF. Validation covers persistent AF in TAILORED-AF, with results for AF lasting six months or more, for women and across NYHA classes, and redo patients with isolated pulmonary veins in the RESTART trial (Heart Rhythm, 2026); a multicenter registry is ongoing.
Use is limited to trained electrophysiologists in hospital labs with named recording and mapping systems. Results for atrial tachycardia and paroxysmal AF are not reported, and pediatric use is not addressed.
Alternatives to Volta Medical
Other products that do the same kind of work as Volta Medical, and what each one does.
Rhythm AI, a UK company with US offices in La Jolla, California, sells the STAR Apollo Mapping System, software that helps electrophysiologists find the sources driving atrial fibrillation while they plan an ablation.
PulseAI is an ECG analysis company in Belfast, Northern Ireland, that licenses deep learning models to read electrocardiograms from smartwatches, ambulatory patches, 12 lead machines and defibrillators, on the device or in the cloud.
Implicity runs a remote monitoring platform for patients with cardiac implantable electronic devices.
AccurKardia makes AccurECG, cloud based software for automated interpretation of ambulatory ECG recordings, sold to cardiac monitoring companies, device manufacturers, hospitals and independent diagnostic testing facilities.
Dilon Medical Technologies makes MarginProbe, an intraoperative device that helps breast surgeons judge, during a lumpectomy, whether cancer remains on the surface of the tissue they have removed, so they can take more tissue before the patient leaves the operating room instead of…
Bayesian Health, a Johns Hopkins spinout founded by Suchi Saria, sells a platform of machine learning models that read the EHR continuously and flag inpatients at rising risk, with workflows that prepare order sets and documentation for clinicians to act on.
Pricing
Vendor-published figures are labeled as such. Figures labeled “Estimated” are derived from third-party sources and have not been confirmed by the vendor.
| Entry Price | Pricing Basis | BAA Tier | Implementation | Source |
|---|---|---|---|---|
|
Not published
|
Sold as a capital system: a dedicated computer with the AF-Xplorer II software, cables, monitor and accessories, installed and configured on site by Volta, with training for electrophysiologists after each installation and software upgrade. No price, license term or per procedure fee is published. | No business associate agreement is offered; the device analyzes signals on a local computer in the EP lab. | Not published. Installation, configuration and operator training are delivered by Volta representatives. | Vendor Published |