Clinical Decision Support
D

Dilon Medical Technologies

Dilon Medical Technologies makes MarginProbe, an intraoperative device that helps breast surgeons judge, during a lumpectomy, whether cancer remains on the surface of the tissue they have removed, so they can take more tissue before the patient leaves the operating room instead of returning for a second surgery. A handheld probe applies radiofrequency fields to each face of the specimen, and a classification algorithm on the console compares the resonant frequency and impedance of each reading against a decision boundary derived from known malignant and benign breast tissue, reporting each margin as positive or negative in roughly five minutes. FDA approved it through premarket approval P110014 in December 2012 as an adjunct to standard methods such as intraoperative imaging and palpation, and FDA lists it among AI enabled devices. In the pivotal randomized trial of 596 patients, surgeons using the device addressed every positive margin during the first operation in 71.8% of patients who had one, against 22.4% with standard care, at a per margin specificity of 45.1%. More than 12 peer reviewed publications cited by Dilon report relative reductions in repeat excision rates of 23% to 88%, and a 2023 study compared its costs with full cavity shave margins. A pivotal trial of a next generation MarginProbe began enrolling in 2024.

Dilon Technologies, Inc., based in Newport News, Virginia, with more than 20 years in medical devices, also sells the TrueView 200 Pro specimen radiography system, made by Compai Healthcare, whose calcification tool boxes suspected calcifications in red; the HEMOBLAST Bellows hemostat; Navigator gamma probes; and the CoPilot VL+ video laryngoscope, in 76 countries. MarginProbe's classifier and TrueView's calcification tool are the company's only AI.

AI Health Index verifiedNovember 23, 2026
Compare Dilon Medical Technologies with other vendors
Founded
—
Headquarters
Newport News, Virginia, United States
Website
dilon.com/
Categories
clinical-decision-support
Indexed Products
MarginProbe System, TrueView 200 Pro specimen radiography, Next Generation MarginProbe
Buyer Segments
Hospital Breast Surgery Programs, Breast Centers, Surgical Oncologists
Assessment

Capability Axes

An AI Health Index grade measures what a buyer can verify from public sources on the date shown. It is not a rating of how good the product is. A vendor can build an excellent system and grade low on an axis because it publishes nothing an outsider can check. How grades read

AI Capability
BB on AI CentralityThe model is the engine of a core module. The platform carries other value, but this capability does not exist without it.
Regulatory Filing

MarginProbe's margin call comes from a fitted classifier. FDA's summary for P110014 says the algorithm labels each probe reading cancer positive or negative by plotting resonant frequency and impedance in a two dimensional space and splitting it with a decision line derived from known malignant and benign breast tissue, one data set choosing the classifiers and a second tuning the margin level threshold.

The radiofrequency sensor, console and single use probe take the measurements, and without the classifier they would return readings with no positive or negative call for the surgeon. Dilon markets the device as radiofrequency technology without calling it AI.

TrueView 200 Pro's calcification tool, which boxes suspected calcifications in red and can be switched off, is one feature of a Compai Healthcare specimen X-ray cabinet, and the hemostat, gamma probes and laryngoscope Dilon also sells work without any model.

BB on Autonomy and Oversight ModelThe oversight structure is described and one part is missing, commonly the threshold at which the system stops or what happens after it is wrong.
Regulatory Filing

MarginProbe is an adjunct: its approved indication pairs it with standard methods such as intraoperative imaging and palpation, and Dilon's page says that when the device identifies cancer on a tissue surface, the surgeon decides whether to remove more tissue.

FDA's summary adds that the device failed to give a reading on 38 of 1,788 margins in the pivotal trial, that malignant and benign readings overlap in its feature space, and that surgeons completed a training phase before randomizing patients. Final pathology still rules on every margin. On TrueView 200 Pro, the operator switches the calcification boxes on or off and reads the image. At the trial's per margin specificity of 45.1%, more than half of the cancer free margins drew a positive call, each one a prompt to take more tissue.

CC on Model and Technology TransparencyThe architecture is described in general terms with nothing identified. Proprietary is asserted rather than explained.
Vendor Published

MarginProbe's page presents the device as radiofrequency electrical fields probing the tissue for remaining cancer, and TrueView's page as advanced radiography with automated software and automatic calcification recognition. The method itself sits in FDA's summary for P110014: resonant frequency and impedance plotted in two dimensions and split by a decision line fitted on training data, running on an embedded processor, with later changes made through PMA supplements.

TrueView 200 Pro's user manual, written by Compai and posted with the product page, says the calcification tool draws a red box around or shades a suspected calcification, and stops at what the tool marks on screen. FDA's database records MarginProbe supplements through 2022, while the next generation device has been in a pivotal trial since enrollment opened in February 2024.

BB on Model Supply Chain DisclosureSubstantial partial disclosure, or a chain that is structurally short: an in house build, a cleared model that cannot be quietly swapped, or a deployment where the transfer does not occur at all. Naming the model provider exits the band below into this one; the axis rises from here on the completeness of the party list and on the terms that govern data once it arrives.
Regulatory Filing

MarginProbe's classifier is Dilon's own, approved with the device under P110014 and run on an embedded processor in the console, so readings are classified in the operating room and the model changes only through a PMA supplement. No cloud service or outside model provider takes part in producing a margin result.

TrueView 200 Pro is a third party device: cleared to Compai Healthcare (Suzhou) under K230136, made by Compai and serviced by Compai's after sales team according to its user manual, so its calcification software comes from Compai rather than Dilon. That manual says patient images travel over a wired or wireless network to the hospital's DICOM server or onto a USB drive, and device data leaves only by USB.

BB on Clinical and Operational EvidenceNamed deployments with dated outcome figures and enough method to test them, or published research short of independent validation.
Vendor Published

MarginProbe's pivotal study was a prospective, multicenter, randomized trial of 596 lumpectomy patients run from 2008 to 2010. Per FDA's summary, surgeons in the device arm addressed every positive margin during the first operation in 71.8% (117 of 163) of patients who had one, against 22.4% (33 of 147) with standard care, though an addressed margin does not mean the final margin was clear.

Dilon's news cites more than 12 peer reviewed publications reporting relative reductions in repeat excision rates of 23% to 88%, by count without a list, and a 2023 paper in Cost Effectiveness and Resource Allocation comparing MarginProbe's cost with full cavity shave margins. The only TrueView claim is the user manual's account of what the calcification tool marks, with no performance figure.

CC on AI Safety and PHI StewardshipGeneral assurances of privacy and security that do not answer the questions artificial intelligence raises: what is retained, what reaches a model, and what happens to it there.
Vendor Published

TrueView 200 Pro's user manual, posted with the product page, describes how patient data moves: images are saved on the system, sent under DICOM 3.0 over the hospital network to its server or copied to a USB drive, and kept behind user logins, access control and audit control. The calcification tool runs on the system itself, so images reach no outside model. MarginProbe classifies readings on an embedded processor in the console in the operating room.

No retention period is set for either device, and nothing says whether images or readings ever return to Dilon or Compai for model development. FDA's summary records the classifier giving no reading on 2% of margins in the pivotal trial.

Regulatory and Compliance
DD on HIPAA and BAA PostureNo statement of status and no privacy document that reaches the product. Both halves are absent at this grade: a service level privacy document on its own lifts a vendor to the band above even where no status is stated. A website privacy notice, including one that says it does not cover customer data, does not reach the product, so a vendor whose only document is that notice and who states no status grades here.
Vendor Published

Patient data sits on both products: TrueView 200 Pro logs in operators, takes patient details from a barcode scanner or DICOM worklist and sends specimen images to PACS, and MarginProbe produces margin readings for each lumpectomy patient.

Dilon makes no statement about HIPAA or business associate status for either product, and dilon.com has no privacy policy at all: the footer lists an address, phone numbers and an email, while the contact forms ask whether the visitor is a patient, provider or hospital representative. The TrueView manual lists health information control among the system's security functions, without a HIPAA reference. Compai's after sales team services TrueView and Dilon's team installs it, with no posted terms covering either team's access to stored patient images.

DD on Security Certifications and Trust CenterControls are asserted with nothing independent behind them, or nothing is published. Read the note before concluding anything: this is the grade most often corrected on a second pass, because assurance material frequently sits on a parent domain or inside an old announcement rather than on the product pages.
Vendor Published

Security material for these devices comes down to one list in the TrueView 200 Pro user manual, written by Compai: least privilege, user access control, audit control, system integrity checks, port access and physical interface locking, external storage media control, backup and restore, and network transmission protection, along with password logins that an administrator resets.

No certification, attestation or penetration test stands behind that list, and the site has no security or trust page. MarginProbe runs on an embedded processor in a standalone console approved in 2012, with no cybersecurity material of its own. The manual's list includes customizing the software update method, without a patch schedule or a route for reporting vulnerabilities in either device.

BB on FDA and Regulatory StatusThe pathway is stated and in progress, or a clearance is named without the vintage and scope a buyer needs to match it to the product on offer. A product outside device regulation also sits here when the vendor publishes what it is for and in what setting and the product on offer matches, or names another regime it works under, such as a registration outside the United States. A general clause saying the service is not medical advice does not reach this band.
Regulatory Filing

MarginProbe holds premarket approval P110014, granted December 27, 2012 to Dilon Medical Technologies, Ltd., with supplements through 2022, for intraoperative assessment of lumpectomy margins as an adjunct to standard methods such as intraoperative imaging and palpation, and FDA lists it among AI enabled devices. Dilon's pages market it inside that use, real time margin information that helps surgeons decide on more tissue, without citing the approval number, date or indication.

TrueView 200 Pro is cleared to Compai Healthcare under K230136 as a cabinet X-ray system. Its user manual lists highlighting suspected calcification among the system's functions, while the 510(k) summary text mentions no algorithm, and Dilon's product page attaches no clearance to the calcification feature it markets.

BB on AI Governance and Bias DisclosureA governance framework with named process behind it, such as certification to an artificial intelligence management standard, or material written for a customer own review committee to evaluate the product with. A framework covering the vendor whole portfolio counts when its scope reaches this product; what it cannot supply is the product specific evaluation the band above asks for. Published results within the one group where the performance of this product is most at risk also sit here: evaluation on the question this axis asks, short of the comparison across groups the band above requires. A medical device clearance or certification, and the quality system behind it, does not reach this band on its own; it is graded under regulatory status. A regulator reviewed plan governing how the model may change does reach it.
Regulatory Filing

FDA's summary for P110014 reports MarginProbe's sensitivity by cancer type from a specimen study of 165 cancerous sites in 50 patients: 0.68 for invasive ductal carcinoma, 0.63 for ductal carcinoma in situ and 0.71 for invasive lobular carcinoma on 7 samples, against 0.70 overall, with confidence intervals for the two larger groups. That table covers sensitivity alone and dates from the 2012 approval.

Results by breast density, patient age, specimen size or race appear in neither the summary nor Dilon's material. Nothing covers how the classifier or TrueView's calcification tool is monitored once installed, and changes to MarginProbe's algorithm go through FDA supplements, with no internal review process of Dilon's own described.

BB on AI Liability and RecourseA published falsifiable commitment, or a real correction route the affected person can exercise against the vendor. A published error rate with its method and denominator grades here. A statutory right that runs to the covered entity rather than to the vendor does not reach this band on its own: every other route here asks something of the vendor, and being located in a particular jurisdiction is not conduct.
Regulatory Filing

FDA's summary for P110014 reports MarginProbe's per margin sensitivity of 73.8% and specificity of 45.1%, with 95% confidence intervals and positive and negative predictive values of 21.6% and 89.4%, across 1,788 margins from the 298 device arm patients of a randomized trial, so the error profile behind each call is on record. The surgeon decides on every margin the device flags, and final pathology later confirms whether margins were clear. No warranty, guarantee or remedy attached to a wrong margin call or a missed calcification is offered for either device.

Integration and Deployment
CC on EHR and Interoperability DepthIntegration is claimed through standards or a middleware layer with no system named and nothing to verify.
Vendor Published

TrueView 200 Pro supports DICOM 3.0 over wired or wireless networks, and its user manual walks through configuring DICOM worklist, MPPS, PACS storage, DICOM print, query and retrieve connections by IP address, AE title and port, with a connection test for each. Patient details come from the worklist or a barcode scanner, and images can also leave on a USB drive. No PACS vendor, EHR or hospital is named where the connection runs in production.

MarginProbe is a standalone console with a single use probe and no stated data connection, so its margin results reach the record through the surgical team's own documentation and the pathology report that follows.

BB on Deployment Model and Data ResidencyOptions and residency are stated with isolation or the processing path left open.
Vendor Published

Both products are equipment installed at the hospital, and neither is described as sending data to Dilon or a cloud service. MarginProbe runs as a console with a sterile single use probe in the operating room, classifying readings on its own embedded processor with no stated data connection.

TrueView 200 Pro is a mobile cabinet X-ray system for a hospital, free standing clinic or breast center, delivered and installed by Dilon's team, and its user manual says images are saved on the system and leave only for the hospital's DICOM server, over a wired or wireless network the hospital configures, or on a USB drive. Where MarginProbe keeps its readings, how long either device stores data and whether service staff reach either device remotely are not covered. Dilon sells in 76 countries, with the same product line presented for every market.

Commercial
DD on Commercial TransparencyNothing a buyer can establish before a sales conversation. A published pricing claim contradicted by evidence also grades here.
Vendor Published

Sales run through a contact form that sorts patients, providers, hospital representatives and investors, and through demo requests for TrueView 200 Pro. No price is given for the MarginProbe console, its single use sterile probes, TrueView 200 Pro or service contracts, and no unit of charge is stated, whether a console purchase, a lease or a fee per probe.

Dilon's 2023 cost study argues that cutting repeat excisions with MarginProbe can save money against full cavity shave margins, a case made without a device price a hospital could weigh against those savings. TrueView is sold as cost effective for any medical setting, also without a figure.

BB on Setting and Specialty CoverageCoverage is named with validation behind part of it.
Vendor Published

Breast conserving surgery is the setting: MarginProbe assesses lumpectomy margins for invasive cancer and ductal carcinoma in situ in the operating room, and TrueView 200 Pro images lumpectomy and core biopsy specimens in hospitals, free standing clinics and breast centers, with a QuadraView specimen holder that captures orthogonal views to measure the distance from a calcification to the cut edge.

The pivotal randomized trial and the peer reviewed publications Dilon cites validate MarginProbe in lumpectomy across multiple centers. TrueView's calcification tool has no validation study behind it, other breast procedures and other cancers sit beyond MarginProbe's indication, and the next generation MarginProbe is still in its pivotal trial.

Alternatives

Alternatives to Dilon Medical Technologies

Other products that do the same kind of work as Dilon Medical Technologies, and what each one does.

MedPearl is a clinical decision platform built inside Providence and in use there since mid 2022.

Veracyte sells genomic tests that pair a laboratory expression assay with a learned classifier, used by urologists, oncologists and endocrinologists to decide how intensively to treat a cancer or whether a thyroid nodule needs surgery.

Heartflow created non invasive coronary physiology from CT.

ArteraAI is the Los Altos, California company at artera.ai that builds multimodal AI cancer tests. It is unrelated to Artera of Santa Barbara (artera.io), which makes patient communication agents.

DermaSensor makes a handheld device that primary care physicians press against a suspicious skin lesion.

Ibex sells AI applications that read H&E whole slide images for prostate, breast and gastric cancer and return their findings into laboratories' existing image management and laboratory systems, through partners including Roche, Philips, Proscia, Sectra and Fujifilm.

Commercial

Pricing

Vendor-published figures are labeled as such. Figures labeled “Estimated” are derived from third-party sources and have not been confirmed by the vendor.

Entry Price Pricing Basis BAA Tier Implementation Source
Not published
MarginProbe is sold as a console used with single use sterile probes, and TrueView 200 Pro as a capital specimen radiography system, both through Dilon's sales team. No price for the console, probes, TrueView or service contracts is published. No business associate agreement or HIPAA position is published. Not published. Dilon's team handles TrueView delivery and installation. Vendor Published

A 2023 study Dilon publicized compares MarginProbe's costs with full cavity shave margins, without a device price.

Contact us

Found a vendor we missed? Have feedback on the index? We’d love to hear from you.