Notal Vision
Notal Vision's SCANLY Home OCT is a prescribed home imaging program for people already diagnosed with neovascular, or wet, age related macular degeneration. The patient scans each eye at home, usually in under a minute, on a spectral domain optical coherence tomography device that captures a 10 by 10 degree area around the point of fixation and sends the images over a built in wireless link to Notal Health Cloud. There the Notal OCT Analyzer, the company's proprietary AI, detects and segments fluid spaces within and beneath the retina, estimates their volume and ranks the scans by fluid area.
The retina specialist reviews annotated scans, en face projections and volume trends in the SCANLY Portal and sets notification thresholds for each eye. FDA granted De Novo authorization under DEN230043 in May 2024 for imaging between office visits, not for treatment decisions. Across two pivotal US trials of more than 500 patients, 96% of participants in the longitudinal study captured images at home, averaging 5.9 scans a week.
The Notal Vision Monitoring Center, led by ophthalmologists and staffed by certified ophthalmic professionals, verifies benefits, enrolls and trains patients, ships the device, tracks adherence and bills insurance, with no fee to the practice. Kester Nahen is chief executive, with US headquarters in Manassas, Virginia and research in Tel Aviv. Notal also runs ForeseeHome, a separate home test for dry AMD.
Capability Axes
An AI Health Index grade measures what a buyer can verify from public sources on the date shown. It is not a rating of how good the product is. A vendor can build an excellent system and grade low on an axis because it publishes nothing an outsider can check. How grades read
The analyzer is what turns daily home scans into monitoring. The SCANLY device captures OCT images and the cloud reconstructs them, and the Notal OCT Analyzer then detects intraretinal and subretinal fluid, segments it, estimates its volume and ranks the scans by fluid area. The physician's notification threshold for each eye is set on that volume, so the alert that brings a patient back before a scheduled injection exists only as model output, and FDA's De Novo authorization covers visualization and volume estimation as one system.
Real value sits around the model: the home OCT hardware, the Monitoring Center's enrollment, training, shipping and adherence work, and the portal. Without the analyzer a practice would receive raw daily scans of every eye with nothing to flag a change.
The physician decides what each eye's alert means. Prescribers set eye specific notification criteria, volume thresholds among them, or choose time based notifications, and every alert goes to the treating retina specialist through the SCANLY Portal. There, scans from different dates sit side by side with the color coded fluid overlay and the signal quality index, and a score below 2 marks a scan as unacceptable.
Labeling keeps the system out of treatment decisions and alongside standard office exams, and the physician responsibilities form assigns interpretation and response to the ordering clinician. What is missing is a route for the model's own errors: a clinician can disagree with a segmentation but has no way to correct it in the portal, and nothing says how a missed or false alert is reported back to Notal.
Two named components produce the output, both Notal's own. The Volume Generator reconstructs the OCT series and computes the manufacturer's signal quality index, and the Notal OCT Analyzer segments intraretinal, subretinal and total fluid on each two dimensional B scan, estimates volumes from the reconstructed series, and displays color coded overlays, en face projections by fluid type and B scans ranked from largest to smallest fluid area.
Schneider and colleagues described the analyzer and tested it against certified reading center graders in Ophthalmology Science in 2024. The size and makeup of its training images, its version history and how an updated analyzer reaches devices already in patients' homes are not published, beyond the upgrade and maintenance plan FDA reviewed with the authorization.
The chain is short and owned by Notal. The home device sends scans over its built in wireless link to Notal Health Cloud, where Notal's own Volume Generator and OCT Analyzer run, and the physician retrieves the results through Notal's SCANLY Portal. No outside model provider sits in the path, and the analyzer is a fixed component of the De Novo authorized system rather than a swappable service.
The parties left unnamed are infrastructure ones: the cloud host behind Notal Health Cloud, the wireless carrier that moves scans from the home, and any analytics or support vendors, along with the terms that govern retinal images once they reach them.
Two prospective pivotal studies, published in Ophthalmology Science with Notal authors and certified reading center graders, carry the evidence. Heier and colleagues (2025) followed 180 participants with wet AMD imaging daily at home for five weeks: they captured 5,426 scans in their study eyes, 98% of attempts succeeded, and masked graders found 86.6% positive and 86.1% negative agreement between home and office images for visible fluid.
Schneider and colleagues (2024) compared the analyzer with three reading center graders. Repeated analyzer volume measurements in eyes above ten volume units varied by 11.1%, against 16.4% for graders reading office scans, with median Dice agreement of 0.61 to 0.68. No practice has reported results from commercial use, such as fewer injections or earlier retreatment.
Retinal scans move from the home device to Notal Health Cloud over a secure connection, are analyzed there and are shown only to prescribing users in the portal. Notal Vision's notice of privacy practices, which it holds as a monitoring provider in its own right, governs the health information it keeps and limits uses to treatment, payment and health care operations, with patient rights of access and amendment.
The notice does not say how long identifiable scans and analyzer outputs are kept, whether de identified images or fluid masks train later analyzer versions, or how home images move into research. For a product built on daily images of every enrolled eye, the stewardship on offer is the general privacy notice rather than terms written for the imaging pipeline.
Notal Vision handles SCANLY data as a health care provider in its own right. Its Monitoring Center bills patients and insurers for the service, its notice of privacy practices covers the health information it holds, disclosure for treatment, payment and operations, and patients' rights to access and amend records, and the SCANLY Portal is described as HIPAA compliant.
What a referring practice does not get is a business associate agreement or a statement of how Notal and the practice divide responsibility for the scans. The physician order and responsibilities forms set clinical duties but no privacy terms, and the subcontractors that handle the images, such as the cloud host, are not addressed.
FDA's review is the outside check. The De Novo submission included software and cybersecurity verification covering unauthorized access and network failures, and the authorized device carries an upgrade and maintenance plan. The home unit sends scans over a secure connection, and the portal limits scans and analyzer results to prescribing users.
That review tests the device's security features rather than how Notal runs its cloud and Monitoring Center. No SOC 2 report, HITRUST certification, ISO 27001 certificate or trust page is published for Notal Health Cloud or the SCANLY Portal, which hold daily retinal images and billing records for every enrolled patient.
SCANLY is the Notal Vision Home Optical Coherence Tomography System, authorized by FDA on May 15, 2024 under De Novo DEN230043 as a Class II prescription device. Its indication covers visualization of intraretinal and subretinal hypo reflective spaces and estimation of their volume in eyes diagnosed with neovascular AMD, imaged at home between regularly scheduled clinic assessments.
Labeling says it is not intended to make treatment decisions or to replace standard office exams, and the marketing stays within that, presenting SCANLY as monitoring between visits with the physician deciding care. Remote OCT codes 0604T to 0606T exist for the service, and some payers cover it claim by claim while broader coverage is worked out.
The authorized labeling names who the device works less well for. In the pivotal cross sectional study, people with vision worse than 20/80 made up a disproportionate share of those who could not calibrate the device, since self imaging depends on holding fixation, and the labeling notes limited clinical performance data for participants of African or Asian descent and for Hispanic or Latino participants.
That is a published result in the group where home imaging is most at risk, poor fixation, with the gaps for the others stated. Analyzer segmentation accuracy is not reported across vision levels, race or ethnicity, fluid type or device generation, and no monitoring of the model's behavior after release is described beyond the maintenance plan.
The portal gives the physician what is needed to challenge a reading. Each scan appears with its color coded segmentation overlay, signal quality index and history, so a volume estimate can be checked against the image and its trajectory before anyone acts. The physician responsibilities acknowledgment puts interpretation, monitoring and response on the ordering clinician, and the warnings and precautions sheet keeps standard office exams in place.
Notal itself takes on no obligation for an analyzer error. The device terms offer no remedy for a missed or false fluid alert, and the amendment right in the privacy notice covers the health record rather than a wrong measurement.
The SCANLY Portal is a separate web workspace. Physicians send prescriptions to the Monitoring Center on Notal's order form, then review annotated scans, clinical reports and volume histories in the portal, where they can view and export data for their own records.
No connection to any record system is named, no standard such as HL7 or FHIR is mentioned, and the portal user manual describes no automatic feed of volumes, images or alerts into the chart. A retina practice runs SCANLY beside its record system and moves anything it wants charted by hand or by export.
Processing happens in one place, Notal Health Cloud. The home device uploads scans over its built in wireless connection, volume reconstruction and the analyzer run in the cloud, and clinicians review results in a browser. There is no local or on premises processing option, and the device ships tied to this service.
Notal is based in Manassas, Virginia, with research and engineering in Tel Aviv, and sells SCANLY in the United States, so US storage is implied. The cloud provider, the country where scans are stored and whether any processing or engineering access crosses to Israel are not stated.
For the practice the cost is stated: nothing. The Monitoring Center supplies the device, verifies benefits, trains patients and bills the patient or insurer for monitoring, and practices pay no fees and store no equipment. Physicians bill their own review under remote OCT codes 0604T, 0605T and 0606T, which let the physician and the Monitoring Center each bill every 30 days, with the coding guidance labeled illustrative.
What the patient or insurer pays is not given as a rate, and coverage is still being worked out with payers, some of which pay claim by claim, so the patient's share depends on the plan.
The coverage is deliberately narrow and studied. SCANLY is for eyes already diagnosed with neovascular AMD, imaged by the patient at home between office assessments under a retina specialist who reviews the results. The five week pivotal study enrolled patients with a mean age of 77 imaging daily, and the analyzer study tested segmentation against certified reading center graders.
Labeling flags where the setting strains, since people with vision worse than 20/80 struggle to calibrate the device. SCANLY is not for screening undiagnosed patients, for dry AMD or for other retinal disease; Notal sells ForeseeHome separately for watching dry AMD for conversion.
Alternatives to Notal Vision
Other products that do the same kind of work as Notal Vision, and what each one does.
ForeseeHome is Notal Vision's prescribed home monitoring program for people with intermediate dry age related macular degeneration at risk of converting to wet AMD, sold under its own brand beside Notal's SCANLY home OCT service.
DentalMonitoring analyzes smartphone images of teeth and orthodontic appliances between office visits, for orthodontists and dental practices in 50 countries.
Dozee, built by Turtle Shell Technologies in Bengaluru, sells contactless patient monitoring to hospitals, skilled nursing facilities and remote care providers.
Neurotrack describes itself as a cognitive care enablement platform: digital cognitive assessments, an adaptive layer it calls Smart Logic, and care planning support, sold to help providers assess, diagnose and care for people with mild cognitive impairment and dementia.
Tyto Care sells remote physical exam devices and software that health systems, health plans and other care organizations run with their own clinicians.
iRhythm runs Zio, a long term ambulatory cardiac monitoring service.
Pricing
Vendor-published figures are labeled as such. Figures labeled “Estimated” are derived from third-party sources and have not been confirmed by the vendor.
| Entry Price | Pricing Basis | BAA Tier | Implementation | Source |
|---|---|---|---|---|
|
$0 for the prescribing practice
$0 baseline
|
The monitoring center bills patients or their insurers for prescribed SCANLY monitoring. Procedure codes 0604T, 0605T and 0606T cover eligible remote OCT services on a 30 day cycle. The referring practice pays no program fee. | Notal Vision bills as a monitoring provider under its own notice of privacy practices, and the SCANLY Portal is described as HIPAA compliant. No business associate agreement is part of physician enrollment. | Notal supplies the device and trains the patient as part of the prescribed monitoring program. There is no practice onboarding charge. | Vendor Published |
Payer coverage is still being worked out, some payers cover the service claim by claim, and the patient's share varies by plan. The practice and the Monitoring Center bill their own services separately, and the device is supplied as part of monitoring with no purchase price.