Radiology & Imaging AI
B

Bot Image

Bot Image, based in Omaha, Nebraska, sells ProstatID, FDA cleared hosted software that helps physicians qualified to read prostate MRI identify regions suspicious for cancer and assess their likelihood of malignancy in screening and diagnostic exams. It analyzes axial T2 weighted, diffusion weighted and apparent diffusion coefficient series from Philips, GE and Siemens scanners at 1.5T and 3T. Deep learning segments the gland, and a random forest trained on biopsy proven examples of normal, benign and cancerous tissue detects and scores suspicious areas. Results come back as color overlays marking each lesion's location, margin and risk, an overall PI RADS suggestion and two 3D views of the gland.

An imaging site sends a study from its PACS to the hosted service, and the output is appended to the patient's study, typically in under five minutes, with no extra hardware or viewer. Sites connect directly or through Blackford and TeraRecon Eureka Clinical AI.

FDA cleared ProstatID under K212783 in July 2022 as a detection and diagnosis aid. In the clearance reader study of 130 cases and nine readers, assisted AUC rose from 0.629 to 0.671, and a peer reviewed 2024 study of 150 exams reported a 0.045 gain. The software is sold per case, and trying it costs nothing.

AI Health Index verifiedNovember 21, 2026
Compare Bot Image with other vendors
Founded
—
Headquarters
Omaha, Nebraska, United States
Website
botimageai.com/
Categories
radiology-and-imaging-ai
Indexed Products
ProstatID
Buyer Segments
Radiology departments, Prostate MRI imaging centers, Urology practices, Medical imaging platform operators
Assessment

Capability Axes

An AI Health Index grade measures what a buyer can verify from public sources on the date shown. It is not a rating of how good the product is. A vendor can build an excellent system and grade low on an axis because it publishes nothing an outsider can check. How grades read

AI Capability
AA on AI CentralityThe artificial intelligence is the product. Remove the model and there is nothing left to sell.
Vendor Published

The models are the product. Deep learning segments the prostate and a random forest estimates a likelihood of malignancy for every suspicious area, and everything ProstatID returns is their output: the color overlays, the per lesion risk scores, the suggested PI RADS assessment and the 3D views of the gland. An imaging site sends the study from its PACS and receives an appended series carrying that reading information.

Remove the models and the DICOM routing, the secure connection and the partner integrations would move images with nothing to add, since the service has no viewer, worklist or reporting function of its own. What a site pays for, case by case, is the model's analysis of that case.

BB on Autonomy and Oversight ModelThe oversight structure is described and one part is missing, commonly the threshold at which the system stops or what happens after it is wrong.
Vendor Published

Processing runs without a click and the physician decides. The service checks the incoming sequences, detects, segments and scores suspicious lesions and appends the results to the exam, and "Results are returned and appended to the unique patient study without any further human interaction." That automation covers processing only.

The radiologist reads the original images beside the overlays, risk scores and suggested PI RADS, and the cleared labeling says patient management decisions should not be made solely on ProstatID and keeps dynamic contrast images, which the device does not analyze, in the reader's hands. No confidence cutoff withholds a weak result, and no route exists for reporting a discordant finding back to Bot Image once a biopsy contradicts the score.

BB on Model and Technology TransparencyThe approach or the suppliers are named without the version and update discipline behind them. Naming a supplier is the entry to this band both here and on Model Supply Chain Disclosure, which ask different questions of the same disclosure: who receives the data, and what produces the output.
Vendor Published

Both learned components are named. The 510(k) summary says "Deep learning and Random Forest algorithms are applied to the DICOM image set of MRI Axial Images", with deep learning segmenting the gland and a random forest classifying patch level features, and the algorithms "are trained with a large database of biopsy-proven examples of normal, benign, and cancerous tissues."

The detection engine takes the axial T2, diffusion and ADC series, finds candidate lesions and assigns each a scaled likelihood score, and the output adds overlays and a whole exam suggestion. The training set carries no case count, the summary's comparison table calls the detection algorithm random forest rather than neural network, and no version history or update schedule accompanies the cleared model.

BB on Model Supply Chain DisclosureSubstantial partial disclosure, or a chain that is structurally short: an in house build, a cleared model that cannot be quietly swapped, or a deployment where the transfer does not occur at all. Naming the model provider exits the band below into this one; the axis rises from here on the completeness of the party list and on the terms that govern data once it arrives.
Vendor Published

The chain is short at the model layer. Bot Image developed the model inside its cleared device, so the analysis cannot be swapped for another supplier's without a new regulatory review, and no outside model provider sits between the image and the score. Distribution adds named parties.

A site connects directly over a PACS to PACS link or routes studies through Blackford or TeraRecon Eureka Clinical AI, and the Precision Prostate Consulting license separates Bot Image from that application provider. The product page also refers to a cloud service provider without naming it, and the compute operator, its location and its retention terms for images in transit differ by route, with nothing written down for any of them.

BB on Clinical and Operational EvidenceNamed deployments with dated outcome figures and enough method to test them, or published research short of independent validation.
Peer Reviewed Publication

Two controlled reader studies test the cleared device. In the 510(k) study of 130 cases with complete follow up and nine readers, reader AUC rose from 0.629 without ProstatID to 0.671 with it (difference 0.042, 95% CI 0.005 to 0.080, p = 0.029), standalone sensitivity was 80% at one false positive per patient and 98% at three, and standalone lesion level AUC was 0.710.

A peer reviewed 2024 study of 150 retrospective exams and nine readers reported a 0.045 gain, with standalone AUC of 0.929 across prostate tissue. The product page's "93.6% AUROC" against a best physician score of 75.9% comes with no method. No result is tied to a named customer site or a software build.

CC on AI Safety and PHI StewardshipGeneral assurances of privacy and security that do not answer the questions artificial intelligence raises: what is retained, what reaches a model, and what happens to it there.
Vendor Published

Data handling is described for the direct route. The processing server takes the required prostate series, deletes the others and returns results to the study's destination, and Bot Image states that its direct service neither processes nor stores protected health information, with studies carried over a secure connection. Training drew on biopsy proven cases and radiologist interpretations.

How long the series it keeps stay on the server, how DICOM header identifiers are handled when whole studies are sent, whether submitted studies may be used to improve the model, and how an inaccurate result is reported and investigated carry no written terms. Studies routed through Blackford and TeraRecon follow those platforms' own handling rules.

Regulatory and Compliance
CC on HIPAA and BAA PostureA status is stated but not supported: compliance is claimed, or business associate status stated, without the underlying document, including where the only privacy notice published covers the website rather than the service that handles patients. The mirror case grades here as well: a substantive privacy document that reaches the service, with no statement of business associate status anywhere and no scope position taken. One of the two elements the band above asks for is present, which is a grade below it and not a grade at the floor.
Vendor Published

A no PHI position is stated without a service document behind it. For the direct hosted route Bot Image says the service processes the required MRI series without processing or storing protected health information, and it calls the connection secure and HIPAA compliant.

The only privacy notice posted covers website visitors, so no service level document explains how identifiers in the DICOM input are removed, held or discarded, and the ProstatID product page itself carries no PHI or HIPAA statement. No business associate agreement is offered for the direct route, and the partner routes fall under Blackford's or TeraRecon's agreements with the site. A written service scope confirming that studies arrive and leave without identifiers would anchor the position.

DD on Security Certifications and Trust CenterControls are asserted with nothing independent behind them, or nothing is published. Read the note before concluding anything: this is the grade most often corrected on a second pass, because assurance material frequently sits on a parent domain or inside an old announcement rather than on the product pages.
Vendor Published

Controls are described with nothing independent behind them. The direct route moves DICOM studies between the site's PACS and the processing server over a secure connection, the server discards series it does not need and returns results to the originating workflow, and the connection is presented as secure and HIPAA compliant. No SOC 2 report, ISO 27001 certificate, penetration test summary or trust center covers the hosted processing, and the website privacy terms reach visitors only.

Imaging platform partners carry their own security programs for studies routed through them, so the direct route rests on Bot Image's description alone. A cloud provider handles the computing, unnamed on the product page.

AA on FDA and Regulatory StatusThe regulatory position is unambiguous and verifiable: a clearance or authorization identifiable in the public databases, with the version and indication it actually covers. A product outside device regulation reaches this band when the vendor publishes why, naming the exclusion it relies on, and markets nothing past it.
Regulatory Filing

ProstatID was cleared on July 8, 2022 under 510(k) K212783 as a radiological computer assisted detection and diagnosis device that identifies regions suspicious for prostate cancer and assesses their likelihood of malignancy, with Transpara 1.6.0, a mammography device, as predicate.

Intended users are physicians qualified to read and interpret prostate MRI, the population is biological males with a prostate gland, and the summary says patient management decisions should not be made solely on ProstatID. The device reads T2 weighted, diffusion and ADC series and does not analyze dynamic contrast enhanced images. Bot Image's marketing for screening, detection and diagnosis stays inside that indication, and "real-time diagnosis" on the product page refers to its five minute turnaround.

DD on AI Governance and Bias DisclosureNothing published on how model behavior is governed or tested. Multilingual operation with no subgroup performance sits here when the vendor markets recognition quality as a strength, because a caller the system failed to understand leaves no complaint and no record.
Vendor Published

Performance across patient and scanner groups goes unreported. ProstatID is trained on biopsy proven normal, benign and cancerous tissue, and the clearance data came from "a sampling of data from various MRI platforms, locations, and physician groups", but the 510(k) summary reports no result by scanner maker, field strength, race, PSA level or gland size. Its only age and PSA analysis tests them as predictors of cancer, not as strata of device accuracy.

Bot Image shows clinical examples from GE and Philips scanners at 1.5T and 3T without measured results. No governance framework, model change plan or bias evaluation accompanies a model sold as "MRI system agnostic" across five hardware configurations.

BB on AI Liability and RecourseA published falsifiable commitment, or a real correction route the affected person can exercise against the vendor. A published error rate with its method and denominator grades here. A statutory right that runs to the covered entity rather than to the vendor does not reach this band on its own: every other route here asks something of the vendor, and being located in a particular jurisdiction is not conduct.
Vendor Published

The license and the clearance data give a buyer something to hold the product to. The end user license on the Precision Prostate Consulting route warrants that the software operates within the user manual's parameters and assigns product and regulatory claims to Bot Image, while disclaiming an error free result, guaranteed accuracy and several categories of damages. Measured error is stated as well.

The 510(k) summary reports 80% sensitivity at one false positive per patient and a lesion level AUC of 0.710, with the cases and method behind them, so a site can test the device against a stated rate. No remedy attaches to an individual missed lesion or false positive, and no commitment covers correcting the model after a discordant biopsy.

Integration and Deployment
BB on EHR and Interoperability DepthNamed systems with read access or one directional writing, or standards support with named deployments behind it.
Vendor Published

Integration happens at the imaging study level. A source PACS sends prostate MRI DICOM studies to the processing server, and ProstatID returns DICOM series, images and reports attached to the original study for reading in the PACS, radiology information system or workstation, so the output lands where the radiologist already reads.

Blackford and TeraRecon Eureka Clinical AI are named distribution integrations, and the device supports five scanner and field strength combinations across GE, Philips and Siemens. The exchange writes into the imaging record and stops there. No connection to an EHR such as Epic carries lesion scores or PI RADS suggestions into the clinical chart or the referring urologist's record.

CC on Deployment Model and Data ResidencyA single hosted option with location implied rather than committed.
Vendor Published

ProstatID is a hosted service with no workstation install. A site connects its PACS to dedicated processing servers, a link the product page says takes "within an hour or so", studies go out over a secure connection, and generated images and reports come back appended to the source study. Sites can also reach the software through Blackford or TeraRecon Eureka Clinical AI, which run the routing on their platforms.

The product page says the analysis runs on "cloud based computing" through a platform partner or cloud service provider without naming either, and the license terms commit no country or region for image processing. No on premises installation is offered for sites that want studies to stay inside their network.

Commercial
CC on Commercial TransparencyNo price is published and the posture is discoverable: a buyer can establish how the product is sold and what drives the cost before contacting the vendor. Most of the index sits here.
Vendor Published

ProstatID is sold per case, directly from Bot Image as a hosted license or through imaging platform partners, and the product page says "it costs NOTHING to try." The license on the Precision Prostate Consulting route permits volume purchasing and group sharing through a common DICOM platform.

Bot Image argues the per case cost "is more than covered by additional billing", pointing to CPT 76377 for eligible three dimensional reconstructions billed with a qualifying imaging code, which is provider reimbursement rather than a price. No per case figure or volume tier is given, so the charge unit is clear while its amount comes through a quote.

BB on Setting and Specialty CoverageCoverage is named with validation behind part of it.
Vendor Published

ProstatID is built for one task in one setting: physicians qualified to read prostate MRI, interpreting screening and diagnostic exams at hospitals and imaging centers on Philips, GE and Siemens scanners at supported field strengths. The clearance reader study drew readers of differing experience from academic and nonacademic settings and data from various platforms and locations, which validates the reading task across that range.

A named radiologist at Independent Imaging Center uses it in practice, and D. Russell Locke, M.D. gives a testimonial on the product page. Results by setting, scanner or reader experience are pooled, so the named coverage is supported as a whole rather than configuration by configuration.

Alternatives

Alternatives to Bot Image

Other products that do the same kind of work as Bot Image, and what each one does.

ANDI, short for Advanced Neuro Diagnostic Imaging, is brain MRI analysis software from Imeka Solutions in Sherbrooke, Quebec.

Quibim, a Valencia company named for quantitative imaging biomarkers in medicine, turns medical images into measurements and predictions and sells AI software as a medical device to hospitals alongside a platform and services for biopharma, after a commercial launch in 2021.

mediaire, a Berlin company founded in 2018, sells the md suite, MRI analysis software that radiology practices and hospital departments install on their own network.

MRIguidance, a Utrecht company founded in 2016 with a US office in New York, sells BoneMRI, deep learning software that turns a dedicated 3D gradient echo MRI sequence into CT like images of bone, so spine and pelvic bone can be assessed from an MRI without the radiation of a CT scan.

Foqus Technologies, a Kitchener, Ontario company cofounded by Dr. Sadegh Raeisi, sells DeepFoqus-Accelerate, AI reconstruction software that lets an MRI department scan the brain up to four times faster.

AISAP sells AI software for cardiac ultrasound at the point of care, built at the Heart Center of Sheba Medical Center in Ramat Gan, Israel.

Commercial

Pricing

Vendor-published figures are labeled as such. Figures labeled “Estimated” are derived from third-party sources and have not been confirmed by the vendor.

Entry Price Pricing Basis BAA Tier Implementation Source
Contact for per case pricing
ProstatID is a hosted software license charged per case, sold directly or through imaging platform partners. The Precision Prostate Consulting license permits volume purchasing and group sharing. Trying the software is free, and the per case rate comes through a quote. For the direct hosted route, Bot Image states that its service processes the requisite MRI series without processing or storing protected health information. A direct connection from the site's PACS to the hosted service is typically set up within about an hour. Implementation costs depend on the route and contract. Vendor Published

Clinical buyers connect to the hosted service or use an imaging platform partner. Eligible three dimensional reconstructions can be billed by the provider under CPT code 76377 alongside a qualifying imaging exam. The case charge for ProstatID is a software expense separate from reimbursement.

Contact us

Found a vendor we missed? Have feedback on the index? We’d love to hear from you.