Diagnostics & Genomics
S

SciBase

SciBase sells Nevisense, a device that analyzes the electrical response of suspicious pigmented skin lesions and returns an adjunctive result for a dermatologist considering a biopsy. The probe applies electrical signals through a disposable electrode at several frequencies and depths. The control unit collects approximately 700 data points and classifies the electrical response. An examination electrode is used for one patient and can support several lesion measurements in the same session.

The device returns a score from 0 to 10. Scores from 0 to 3 are classified as negative and scores from 4 to 10 as positive. A dermatologist who has completed Nevisense training initiates the examination and interprets the result with clinical and historical findings. A trained healthcare professional can take the measurement. The current US workflow includes device checks for measurement quality and no longer requires the operator to take a separate reference measurement.

The technology grew out of research at Karolinska Institutet in Stockholm, and FDA approved the device for melanoma in 2017. SciBase manufactures the system and sells it with recurring orders of examination electrodes. It counts 500,000 evaluations in more than 500 clinics worldwide, and Palm Beach Dermatology Group took delivery of six units with electrodes in 2025. The examination result is shown on the device's display for review during the visit. In Europe the device is also CE marked for keratinocyte cancers.

AI Health Index verifiedNovember 25, 2026
Compare SciBase with other vendors
Founded
—
Headquarters
Sundbyberg, Sweden
Categories
diagnostics-and-genomics
Indexed Products
Nevisense, Nevisense examination electrodes
Buyer Segments
Dermatology practices, Dermatology departments, Skin cancer clinics
Assessment

Capability Axes

An AI Health Index grade measures what a buyer can verify from public sources on the date shown. It is not a rating of how good the product is. A vendor can build an excellent system and grade low on an axis because it publishes nothing an outsider can check. How grades read

AI Capability
BB on AI CentralityThe model is the engine of a core module. The platform carries other value, but this capability does not exist without it.
Regulatory Filing

The Nevisense melanoma classifier turns lesion impedance measurements into a score a dermatologist can use during a biopsy assessment. Its original model was fitted using electrical measurements from nearly 2,000 patients across European development sites, primarily from the second International Melanoma Algorithm Training Study. The current workflow uses electronic qualification and feedback on the lesion measurement.

The hardware collects the impedance spectra independently of the risk classification, and the broader Nevisense platform uses those measurements for other clinical and research applications. The learned classifier drives the melanoma assessment.

AA on Autonomy and Oversight ModelWhat the system may do and what it may not do are both published, with escalation thresholds, override paths and the conditions that route a case to a person.
Regulatory Filing

The melanoma score is negative from 0 to 3 and positive from 4 to 10. The device checks measurement qualification and provides lesion feedback when it detects a potential operator error. A healthcare professional who completed device training can perform the examination after a trained dermatologist initiates it. The dermatologist interprets every result used for the biopsy decision alongside the patient's clinical and historical findings. The indication excludes clinically obvious melanoma and lesions a clinician has already decided to biopsy. The physician retains the decision even when the score is negative.

BB on Model and Technology TransparencyThe approach or the suppliers are named without the version and update discipline behind them. Naming a supplier is the entry to this band both here and on Model Supply Chain Disclosure, which ask different questions of the same disclosure: who receives the data, and what produces the output.
Regulatory Filing

The device takes impedance measurements at several frequencies and depths, about 700 data points an exam, and a proprietary classifier that SciBase calls AI turns them into the score. The classifier was developed from nearly 2,000 patients at more than 19 European sites, with the IMATS II dataset as the primary development set. FDA authorized the 2020 change that removed the reference measurement and added an in device qualification module as a supplement to the original approval. SciBase does not give the current classifier's revision or say whether it was refitted when the measurement procedure changed.

BB on Model Supply Chain DisclosureSubstantial partial disclosure, or a chain that is structurally short: an in house build, a cleared model that cannot be quietly swapped, or a deployment where the transfer does not occur at all. Naming the model provider exits the band below into this one; the axis rises from here on the completeness of the party list and on the terms that govern data once it arrives.
Regulatory Filing

SciBase manufactures the diagnostic system and developed the lesion classifier itself from clinical measurement datasets, so no outside model provider is involved. The control unit calculates the result at the practice. An FDA approved manufacturing change identifies Kitron as contract manufacturer of components that include the probe and control unit. Examination data leaves the device only when practice staff export the records.

CC on Clinical and Operational EvidenceNamed customers, or vendor reported percentages with no method, denominator or reference standard. Scale of use is recorded here and is not treated as evidence of benefit.
Regulatory Filing

Palm Beach Dermatology Group took delivery of six Nevisense devices and examination electrodes for clinical use in 2025, and SciBase counts 500,000 evaluations in more than 500 clinics worldwide. The original authorization rested on a blinded pivotal study of 1,951 patients that compared the device assessment with histology, with 95.9% sensitivity and 31.1% specificity in FDA's primary effectiveness analysis.

A 2020 reader study reported 14% more melanomas detected when clinicians had the score. The pivotal study ran on the workflow before the 2020 software change that removed the reference measurement and added qualification checks, and no result series from the current workflow at a named practice has followed.

CC on AI Safety and PHI StewardshipGeneral assurances of privacy and security that do not answer the questions artificial intelligence raises: what is retained, what reaches a model, and what happens to it there.
Vendor Published

The examination unit processes the electrical measurements and displays the classification in the clinic. It keeps a local archive and offers removable media for transferring examination data. The original classifier was trained on dedicated clinical study data. The current device screens the quality of each lesion measurement at the unit and gives feedback on operator errors. SciBase sets no retention or reuse rules for examination data that a practice exports to the manufacturer or to a research workflow.

Regulatory and Compliance
DD on HIPAA and BAA PostureNo statement of status and no privacy document that reaches the product. Both halves are absent at this grade: a service level privacy document on its own lifts a vendor to the band above even where no status is stated. A website privacy notice, including one that says it does not cover customer data, does not reach the product, so a vendor whose only document is that notice and who states no status grades here.
Vendor Published

The dermatology practice takes the lesion measurement and views the assessment on the control unit. Examination records can be archived or exported through the device's removable media workflow, and the practice enters the result in the patient's chart itself. SciBase's privacy terms reach account and ordering information and use of its website and app, and stop short of examination records. SciBase states no business associate status for US practices and offers no BAA for support or research work that would expose identifiable examination records.

DD on Security Certifications and Trust CenterControls are asserted with nothing independent behind them, or nothing is published. Read the note before concluding anything: this is the grade most often corrected on a second pass, because assurance material frequently sits on a parent domain or inside an old announcement rather than on the product pages.
Vendor Published

Nevisense runs on a dedicated control unit with a trained operator and a local output. The software qualifies the measurement and gives feedback when it detects a likely operator error. The device and disposable electrodes are made under regulated manufacturing and quality checks, including approved changes to how finished electrodes are tested. SciBase cites no independent security assessment of the device, its exported files or any support access to them.

BB on FDA and Regulatory StatusThe pathway is stated and in progress, or a clearance is named without the vintage and scope a buyer needs to match it to the product on offer. A product outside device regulation also sits here when the vendor publishes what it is for and in what setting and the product on offer matches, or names another regime it works under, such as a registration outside the United States. A general clause saying the service is not medical advice does not reach this band.
Regulatory Filing

FDA approved Nevisense under PMA P150046 in June 2017 as an adjunct on cutaneous lesions with clinical or historical characteristics of melanoma, when a dermatologist wants more information before a biopsy decision. The approval bars use on clinically obvious melanoma and limits it to intact primary lesions of 2 to 20 mm away from acral, genital, eye and mucosal sites. Supplement S004 in April 2020 removed the reference measurement and added a qualification and feedback module.

S005 in February 2026 let other trained healthcare professionals operate the device while reserving initiation and interpretation for a trained dermatologist. The US pages sell the device as “FDA-approved for melanoma detection” and promise to “Rule out melanoma with 99% NPV” without the PMA number or the adjunctive indication. Keratinocyte cancer indications are CE marked only, and skin barrier uses are sold for research.

CC on AI Governance and Bias DisclosureResponsible artificial intelligence is committed to in policy language with no evaluation behind it. Most of the index sits here.
Regulatory Filing

The original melanoma classifier was evaluated against histology, with results reported by age, sex and lesion characteristics. Sensitivity was notably lower in the youngest group, and the study held too few melanomas in Fitzpatrick skin types V and VI to verify sensitivity there. Those results come from the workflow before the reference measurement was removed. The current software qualifies each lesion measurement and gives feedback on likely operator errors, and no comparison across patient groups has been reported for it.

CC on AI Liability and RecourseMechanisms exist that let someone challenge an output, such as audit trails, source traceability or review before commit, with nothing standing behind the output and no route for the harmed party.
Regulatory Filing

The dermatologist reviews the score with the clinical findings before deciding whether to biopsy. The qualification checks and lesion feedback let the clinical user catch a measurement problem before reading the result. The original device trial supplies an error profile, including specificity of 31.1% against histology, measured before the 2020 change to the measurement procedure. A disputed result is resolved in the clinician's review, and SciBase offers no warranty or correction remedy for a wrong score ahead of the purchase terms.

Integration and Deployment
DD on EHR and Interoperability DepthNo integration evidence. A connector described as available on request grades here until one exists.
Regulatory Filing

The control unit presents the assessment on its own touchscreen. A dermatologist reviews it alongside examination and dermoscopy findings during the visit. The device supports local export to removable media for archiving or further analysis, and the research workflow exports quantitative measurements for tracking. These are device and file workflows, so entering a result in the patient's chart is a manual step for the clinic. Nevisense has no interface to a record system.

BB on Deployment Model and Data ResidencyOptions and residency are stated with isolation or the processing path left open.
Regulatory Filing

The control unit runs the analysis in the dermatology office and presents the classification at the device. Its examination workflow takes the probe measurement and performs the analysis at the same unit. Local archiving or export to removable media enables practices to keep their own copy of the result. The US device operates in the office with a dedicated probe and trained users. The deployment keeps processing at the device, while the practice governs where its exported records are stored.

Commercial
BB on Commercial TransparencyA price or a pricing basis is published without full tiers, so a buyer can size the cost before making contact.
Vendor Published

Nevisense is sold as a device purchase with recurring disposable electrode orders. SciBase's 2025 US price was $7,500 to $10,500 a device, and melanoma examination electrodes cost about $69 to $73 each. An electrode serves one patient and supports as many as 20 measurements in that patient's session, so a practice's running cost follows its exam volume. Orders go through SciBase's US sales team. A patient self pay model, introduced for the US, lets a practice offer the test where insurance does not cover it.

BB on Setting and Specialty CoverageCoverage is named with validation behind part of it.
Vendor Published

The intended US setting is dermatology assessment of cutaneous lesions with clinical or historical features of melanoma when the dermatologist is considering a biopsy. Use is limited to intact primary lesions of 2 to 20 mm, away from acral, genital, eye and mucosal sites. The measurement fits within an office visit, and the trained dermatologist who initiates the examination weighs the result alongside dermoscopy and other clinical findings.

Palm Beach Dermatology Group received six devices for its offices. Historical validation covered multiple centers, and the device's wider markets include skin barrier research and European indications for keratinocyte cancers under a CE mark. No result series has been reported from the current workflow in a named US dermatology program.

Alternatives

Alternatives to SciBase

Other products that do the same kind of work as SciBase, and what each one does.

DermaSensor makes a handheld device that primary care physicians press against a suspicious skin lesion.

Skin Analytics makes DERM (Deep Ensemble for the Recognition of Malignancy), an AI medical device that assesses dermoscopic images of skin lesions for cancer and, in its pathways, discharges low risk patients on its own while fast tracking higher risk ones to a specialist.

CorVista Health sells a cardiopulmonary assessment that combines a resting signal capture with three machine learned analyses.

Stratipath Breast assigns early stage invasive breast cancers to low or high risk groups from routine H&E slides, as a faster, cheaper alternative to gene expression tests for pathology laboratories and breast cancer teams in Europe.

B-Secur licenses HeartKey, a family of electrocardiogram analysis libraries for medical device companies and cardiac monitoring platforms.

Foundation Medicine, based in Boston and a member of the Roche Group, runs comprehensive genomic profiling for cancer in its own laboratories.

Commercial

Pricing

Vendor-published figures are labeled as such. Figures labeled “Estimated” are derived from third-party sources and have not been confirmed by the vendor.

Entry Price Pricing Basis BAA Tier Implementation Source
$7,500 to $10,500 per device
$7,500 baseline
For 2025, SciBase's US Nevisense device price was $7,500 to $10,500. Disposable melanoma examination electrodes cost about $69 to $73 each in the US, with one electrode per patient. The office device processes the exam locally. SciBase offers no BAA for manufacturer access to patient records. Device operation requires completed Nevisense training. The device is purchased separately from the recurring order of examination electrodes. Vendor Published

An electrode covers up to 20 measurements in one patient's session. The device and supplies are purchased through SciBase. A US self pay model lets practices offer the test to patients without coverage.

Contact us