The Accuryn Monitoring System is a bedside fluid measurement system originally submitted by Potrero Medical and now sold by Accuryn Medical of Hayward, California. It combines a SmartFoley catheter, a collection set and a portable monitor. The monitor measures urine volume, core temperature and, with the pressure sensing catheter, intra abdominal pressure. Active Drain Line Clearance moves urine through the line into the measurement cassette to reduce retention and airlocks. The SmartFoley system includes valves that limit backflow and an insertion tray with a catheter fixation device.

AccuTab shows trends for the three measurements, lets staff annotate interventions and displays when urine output meets KDIGO criteria for acute kidney injury. Hospitals can send the measurements to an electronic medical record flow sheet. The FDA cleared the original monitoring system in 2016 and added 12 and 14 French SmartFoley tray configurations in 2023.

Regional One Health’s Firefighters Burn Center tracked 2,243 patients across the years before and after adoption and recorded 40 catheter associated infections before implementation and two afterward. A separate 2024 study across nine tertiary centers compared urine output alerts with serum creatinine in 522 cardiac surgery patients. Stage 1 or higher urine output alerts led the stage 2 or higher creatinine criterion by a median of 33.6 hours.

AI Health Index verifiedNovember 24, 2026
Compare Accuryn with other vendors
Founded
—
Headquarters
Hayward, CA
Categories
inpatient-monitoring
Indexed Products
Accuryn Monitoring System, Accuryn Monitor, SmartFoley, AccuTab, Active Drain Line Clearance, KDIGO Alerts
Buyer Segments
Acute Care Hospitals, Intensive Care Units, Cardiac Surgery Services, Burn Centers
Assessment

Capability Axes

An AI Health Index grade measures what a buyer can verify from public sources on the date shown. It is not a rating of how good the product is. A vendor can build an excellent system and grade low on an axis because it publishes nothing an outsider can check. How grades read

AI Capability
DD on AI CentralityNo learned component can be found. Either artificial intelligence is claimed in the marketing and cannot be located in the product, or the term is covering rules and automation that predate it. The note says which. A product that makes no AI claim and runs on rules, formulas or human curation is outside the index.
Regulatory Filing

The Accuryn Monitor measures urine volume in a cassette, pressure through the catheter's pressure lumen and temperature through its thermistor. Active Drain Line Clearance moves urine through the collection line to the cassette. The kidney alerts compare the recorded urine output with KDIGO criteria, and AccuTab shows when the oliguria criteria are met.

Potrero Medical developed a distinct machine learning system called AKI Predict and received a Breakthrough Device Designation for that system in 2022. The monitor sold today alerts on measured urine output against those criteria, so the kidney alerts marketed as "predictive" contain no learned component.

DD on Autonomy and Oversight ModelNo oversight structure is published. An absolute claim that the system does not err grades here too, because a buyer who believes it will not build the review step that would catch a failure.
Vendor Published

The bedside monitor measures urine and displays pressure and temperature continuously. AccuTab shows the time series, alerts the clinical team when urine output reaches KDIGO criteria and allows staff to record an intervention alongside the trend. The drainage mechanism clears the collection line automatically, and Active Drain Line Clearance is sold with the line "No retained urine. No manual milking. No wrong numbers."

The monitor sends its readings to the flow sheet when integrated, without a separate sign off on each measurement. Clinical staff decide how to interpret the values and whether to intervene. The trend window aligns the alert with earlier urine output and annotated interventions.

BB on Model and Technology TransparencyThe approach or the suppliers are named without the version and update discipline behind them. Naming a supplier is the entry to this band both here and on Model Supply Chain Disclosure, which ask different questions of the same disclosure: who receives the data, and what produces the output.
Regulatory Filing

The SmartFoley uses a thermistor for core temperature and a pressure lumen to measure intra abdominal pressure. Urine passes from the catheter into a measurement cassette, and the monitor displays the measured volume. Active drainage clears retained liquid from the line before measurement. AccuTab compares the volume trend with KDIGO oliguria criteria and displays the relevant stage when those criteria are met.

Clinical staff can annotate events on the trend display to see how readings changed. The monitor and its software are described at the level of sensing and threshold logic, with no learned parameters in the bedside alert flow.

BB on Model Supply Chain DisclosureSubstantial partial disclosure, or a chain that is structurally short: an in house build, a cleared model that cannot be quietly swapped, or a deployment where the transfer does not occur at all. Naming the model provider exits the band below into this one; the axis rises from here on the completeness of the party list and on the terms that govern data once it arrives.
Regulatory Filing

Potrero Medical filed the original Accuryn system with the FDA. Accuryn Medical now sells the monitor, collection sets, connected SmartFoley configurations and the AccuTab display under the Accuryn name. The urine cassette, temperature sensor and pressure lumen send readings to the Accuryn Monitor and its local software. The monitor can operate from the bedside even before the hospital connects its electronic flow sheet.

The value that reaches the flow sheet is calculated in the monitor from the connected sensors. The current workflow uses Accuryn's own monitor and disposables, with the medical record as the optional downstream destination.

AA on Clinical and Operational EvidencePeer reviewed or independently evaluated performance, prospective and multi site where the claim requires it, with the method available to read. The grade describes the strength of the evidence, not its direction: a rigorous study that found no benefit grades by its design, and the note leads with what it found.
Peer Reviewed Publication

Regional One Health's Firefighters Burn Center recorded 40 catheter associated infections among 1,317 patients before implementation and two among 926 afterward, in a retrospective study comparing 2015 through 2018 with January 2020 through April 2022. The charts did not reliably show which catheter each patient received, so the figures describe the unit's adoption periods.

A 2024 prospective and retrospective observational study in the Journal of Cardiothoracic and Vascular Anesthesia, drawn from the Accuryn Registry with an Accuryn employee among its authors, analyzed 522 cardiac surgery patients at nine tertiary centers. Urine output alerts at stage 1 or higher led creatinine at stage 2 or higher by a median of 33.6 hours, and stage 1 alerts were associated with longer ICU and hospital stays and continuous renal replacement therapy.

CC on AI Safety and PHI StewardshipGeneral assurances of privacy and security that do not answer the questions artificial intelligence raises: what is retained, what reaches a model, and what happens to it there.
Vendor Published

The monitor receives temperature, pressure and urine output at the bedside. AccuTab can display those trends elsewhere in the unit and the medical record can receive the measurements in its flow sheet. Potrero Medical’s privacy terms apply to use of its products and allow collection of patient outcomes for research and statistical purposes. It permits distribution of product and service statistics in aggregate after deidentification.

General retention follows legal obligations and the purpose for which the data was collected. Those provisions do not set a retention period for bedside measurements or distinguish clinical research use from model development.

Regulatory and Compliance
CC on HIPAA and BAA PostureA status is stated but not supported: compliance is claimed, or business associate status stated, without the underlying document, including where the only privacy notice published covers the website rather than the service that handles patients. The mirror case grades here as well: a substantive privacy document that reaches the service, with no statement of business associate status anywhere and no scope position taken. One of the two elements the band above asks for is present, which is a grade below it and not a grade at the floor.
Vendor Published

Potrero Medical applies its privacy policy to people who use its products and services, as well as to purchasers, patients involved in claims or investigations and participants whose outcomes are collected for research. Its terms describe administrative, technical and physical safeguards and written assurances for parties receiving information across borders. Accuryn’s bedside monitor can send its measurements to a hospital record when the hospital connects it. The company handles privacy requests directly. A hospital’s use of the connected flow sheet is not paired with a stated business associate status or BAA tier for the monitoring system.

CC on Security Certifications and Trust CenterControls are described with an outside check behind them, such as independent penetration testing on a stated cadence, but no attestation against a recognized framework. A certification that tests the software security features rather than the vendor operations, such as federal health IT certification, is an outside check of this kind.
Regulatory Filing

The current monitor and SmartFoley configurations went through software verification and validation and a cybersecurity risk assessment for the 2023 FDA clearance, which added the 12 and 14 French pressure sensing trays. That testing references the FDA's medical device cybersecurity guidance and AAMI TIR57, along with software lifecycle verification under IEC 62304. The review covers the security of the device itself, not how Accuryn runs its own operations. The company applies physical, administrative and technical safeguards to the personal information it handles. No SOC 2, ISO 27001 or HITRUST attestation is named, and no trust center is offered.

BB on FDA and Regulatory StatusThe pathway is stated and in progress, or a clearance is named without the vintage and scope a buyer needs to match it to the product on offer. A product outside device regulation also sits here when the vendor publishes what it is for and in what setting and the product on offer matches, or names another regime it works under, such as a registration outside the United States. A general clause saying the service is not medical advice does not reach this band.
Regulatory Filing

Accuryn monitors urine output and core temperature and, with the pressure sensing SmartFoley, intra abdominal pressure. FDA clearance K153655 covers the original monitor and K221020 adds the 12 and 14 French pressure sensing tray systems. The pressure measurement can aid in identifying intra abdominal hypertension and abdominal compartment syndrome.

Accuryn markets predictive kidney alerts beside its clearance badge and says it “detected Stage 2 AKI an average of 33.6 hours before serum creatinine.” The study measured the median lead from stage 1 or higher urine output alerts. A separate machine learning algorithm called AKI Predict received Breakthrough Device Designation in 2022. The 2023 clearance covers the sensing and monitoring indications, while its marketed kidney alert runs on KDIGO urine output criteria.

DD on AI Governance and Bias DisclosureNothing published on how model behavior is governed or tested. Multilingual operation with no subgroup performance sits here when the vendor markets recognition quality as a strength, because a caller the system failed to understand leaves no complaint and no record.
Vendor Published

The bedside alert uses urine output and the KDIGO oliguria criteria. AccuTab shows the trend for pressure, volume and temperature alongside events a clinician annotates. The 2024 cardiac surgery study reports how a stage 1 or higher alert compares with a later stage 2 or higher creatinine criterion across 522 patients, and the monitor itself went through validation for its pressure sensing configuration. Those studies describe specific clinical comparisons. The marketed alert does not show how its lead time or false alerts vary by patient subgroup, and the device applies the same threshold framework across the stated settings.

BB on AI Liability and RecourseA published falsifiable commitment, or a real correction route the affected person can exercise against the vendor. A published error rate with its method and denominator grades here. A statutory right that runs to the covered entity rather than to the vendor does not reach this band on its own: every other route here asks something of the vendor, and being located in a particular jurisdiction is not conduct.
Peer Reviewed Publication

Each alert can be compared with the urine output trend in AccuTab. Staff can annotate an intervention and observe the readings that follow. In the 2024 cardiac surgery study, the urine output alert had a discriminative value of 0.78 against creatinine defined stage 2 or higher in a cohort of 522 patients. The nine center study also compared alerts with hospital length of stay, renal replacement therapy and 30 day mortality. The device terms provide a quality contact for instructions and product concerns. The study and the service contact do not establish a reimbursement or remedy when a reading is incorrect.

Integration and Deployment
CC on EHR and Interoperability DepthIntegration is claimed through standards or a middleware layer with no system named and nothing to verify.
Vendor Published

The Accuryn monitor sends measured urine output, pressure and temperature to an electronic medical record flow sheet when the hospital connects it, putting the readings in the record the clinical team uses to document the patient's course. Values reach the flow sheet as measured, with no nurse sign off on each reading. AccuTab also provides a local display of the measurements, trend lines and intervention annotations for nurses and physicians. The monitor can start measuring before the hospital connects its flow sheet. No record system, interface standard or version is named for that connection, and no hospital is named as running it.

BB on Deployment Model and Data ResidencyOptions and residency are stated with isolation or the processing path left open.
Vendor Published

The portable Accuryn Monitor sits at the hospital bedside. It measures the urine collected in the cassette and receives temperature and pressure directly from the catheter. AccuTab can mount to an IV pole and show the readings away from the bedside, including a hall or a location outside the sterile field. Hospitals can connect the monitor to their electronic flow sheet, and the monitor starts operating before that integration is configured. The deployment has a local measurement path. Its privacy terms permit transfers of some personal information to affiliates and service providers under written assurances.

Commercial
DD on Commercial TransparencyNothing a buyer can establish before a sales conversation. A published pricing claim contradicted by evidence also grades here.
Vendor Published

Accuryn sells the monitoring system to hospitals through a quote and a bedside demonstration with a clinical specialist. A configuration pairs a reusable monitor with collection sets and SmartFoley catheters, plus the AccuTab display and a catheter adapter option. A hospital can start with one unit before expanding to more departments, and the flow sheet connection is set up separately. The purchase case rests on acute kidney injury, which CMS now classifies as a hospital acquired harm. What the hospital pays for the monitor, the disposable trays and the integration comes with the quote, with no unit or tier stated beforehand.

BB on Setting and Specialty CoverageCoverage is named with validation behind part of it.
Peer Reviewed Publication

The monitor serves inpatient teams caring for catheterized patients who need continuous fluid, temperature or pressure measurements, in surgical ICUs, cardiac surgery, burn care, trauma, pediatrics and postpartum units. The installed base covers more than 70 US hospitals, among them six children's hospitals, 19 teaching hospitals and 13 VA or Department of Defense facilities.

The Firefighters Burn Center at Regional One Health measured infection incidence across 2,243 admissions before and after its rollout, and nine tertiary centers evaluated the urine output alert in 522 cardiac surgery patients. Postpartum, pediatric and trauma use have no comparable published outcome.

Alternatives

Alternatives to Accuryn

Other products that do the same kind of work as Accuryn, and what each one does.

Biobeat sells wearable vital sign monitors and a cloud platform for hospitals, long term care and home programs.

Bayesian Health, a Johns Hopkins spinout founded by Suchi Saria, sells a platform of machine learning models that read the EHR continuously and flag inpatients at rising risk, with workflows that prepare order sets and documentation for clinicians to act on.

Sickbay, from Medical Informatics Corp (MIC) in Houston, is a vendor neutral platform that captures physiologic data and waveforms from bedside monitors, ventilators and other devices across manufacturers, stores them, and displays them in a browser for bedside teams, telemetry…

AITRICS is a Seoul medical AI company whose AITRICS-VC (VitalCare) predicts deterioration in hospitalized adults from routine electronic medical record data.

Dozee, built by Turtle Shell Technologies in Bengaluru, sells contactless patient monitoring to hospitals, skilled nursing facilities and remote care providers.

VSTOne is VirtuSense's acute care product, a single sensor device per hospital room that predicts when a patient is about to leave the bed or chair unassisted and alerts staff before it happens.

Commercial

Pricing

Vendor-published figures are labeled as such. Figures labeled “Estimated” are derived from third-party sources and have not been confirmed by the vendor.

Entry Price Pricing Basis BAA Tier Implementation Source
Hospital quote
Accuryn sells the bedside monitor and SmartFoley configurations through a hospital quote. Clinical specialists demonstrate the device in the unit. Potrero Medical applies its privacy terms to product use and collection of patient outcomes for research. Flow sheet integration is optional. The bedside monitor starts taking measurements ahead of that connection. Vendor Published

The quote covers a configuration of the monitor, catheter and collection equipment. The purchase case rests on acute kidney injury, which CMS now classifies as a hospital acquired harm.

Contact us

Found a vendor we missed? Have feedback on the index? We’d love to hear from you.