4DMedical
4DMedical, based in Carlton, Australia with US offices in Woodland Hills, California, sells pulmonary imaging software as a service to hospitals, radiology practices and lung disease centers. CT:VQ turns paired noncontrast chest CTs into ventilation and perfusion maps registered to the anatomy, with lung volumes, lobar distribution, DICOM structured reports to PACS and a patient facing report, and FDA cleared it under K251484 in August 2025. CT LVAS measures regional ventilation from paired inspiratory and expiratory CTs, XV LVAS measures airflow from fluoroscopy, and the portfolio adds lung density, coronary artery calcification and pulmonary hypertension tools; lung texture analysis is for clinical investigation only in the United States.
The AI sits in IQ UIP, which came with the 2023 Imbio acquisition. Its locked deep learning classifier reads qualifying adult chest CTs and puts findings suggestive of usual interstitial pneumonia on a separate worklist for specialists at interstitial lung disease centers, while the study stays in PACS for the thoracic radiologist's read. FDA cleared it under K242467 in December 2024; on 804 held out exams it reached 90.2% sensitivity and 91.5% specificity, with results reported by scanner maker, kernel and slice thickness.
CT:VQ runs at UC San Diego Health, Cleveland Clinic and Stanford, and its perfusion measures tracked lung diffusing capacity more closely than SPECT in its submission data. Scans can be billed under Category III CPT codes, and software is priced by contract, with XV Technology charged per scan.
Capability Axes
An AI Health Index grade measures what a buyer can verify from public sources on the date shown. It is not a rating of how good the product is. A vendor can build an excellent system and grade low on an axis because it publishes nothing an outsider can check. How grades read
The learned model is IQ UIP, a locked deep learning classifier that flags chest CTs suggestive of usual interstitial pneumonia for a specialist worklist, which 4DMedical gained with Imbio and markets as early IPF detection.
The company's lead products work without it. CT:VQ is described as noncontrast post processing that registers inspiratory and expiratory scans into ventilation and perfusion maps, and CT LVAS and XV LVAS measure regional ventilation and airflow, with no learned component claimed for any of them; lung density, calcification and pulmonary hypertension tools sit alongside. Remove IQ UIP and 4DMedical still sells functional lung imaging to the same departments, without the referral alert.
IQ UIP's automated action and its limits are written into the labeling. The locked model flags a qualifying scan for a separate specialist worklist when its score passes 0.395, rejects invalid inputs such as studies with more than one series or no detectable lungs, and leaves the original CT untouched in PACS, where a thoracic radiologist makes the final call with the rest of the patient's information. The indication limits it to notification, not diagnosis.
What happens after an error is not described. Nothing gives the radiologist a way to record a missed or spurious flag in the product, and no follow up of unflagged scans or feedback to 4DMedical is set out.
The model is described in its labeling. IQ UIP is a locked deep learning inference model that classifies a qualifying inspiratory chest CT, developed on anonymized retrospective volumetric CT from hospitals, imaging centers and databases, with development and evaluation sets separated by site and patient where site data allowed and the 804 exam performance cohort quarantined from tuning. It applies a binary threshold of 0.395 and states acquisition constraints and known limitations.
The network architecture is not named, and because the model is locked, any change would come as a new release under FDA review; no version history or schedule for one is published. CT:VQ's processing method is not described.
The model chain is short because the model is in house and locked. Imbio developed and manufactures IQ UIP, 4DMedical sells it in its pulmonary portfolio, and the cleared classifier runs at a fixed operating point that cannot be swapped without a new submission. Images enter through the provider's imaging workflow, and Laurel Bridge Software and Kailo Medical carry the packaged exchange.
The infrastructure layer is not listed. No cloud host or subprocessor list is published for the 4DMedical platform, distribution partners can also run IQ UIP under their own arrangements, and the data terms that apply on those partner routes are not stated.
Regulatory validation carries the evidence. For IQ UIP, the K242467 submission tested 804 chest CT exams held apart from tuning against a thoracic radiologist reference, and at the 0.395 threshold sensitivity was 90.2% and specificity 91.5%. For CT:VQ, submission data on file show its perfusion heterogeneity measures correlating with lung diffusing capacity at r squared 0.557 against 0.380 for SPECT, and a reader study found agreement with SPECT V/Q across lung zones at Kendall's tau of 0.702 to 0.765, above a target of 0.40.
UC San Diego Health, Cleveland Clinic and Stanford are named CT:VQ sites without reported outcomes, and no site reports results from routine IQ UIP use, such as earlier referral or diagnosis.
The design pushes identity handling to the customer. 4DMedical's terms make the imaging provider responsible for de identifying images before sending them and re identifying returned results, with Laurel Bridge Software and Kailo Medical supplying the packaged transfer, and Imbio's business associate agreement limits use of protected information, carries those limits to subcontractors and requires return or destruction at the end of the contract where feasible.
No retention period is given for submitted scans, derived maps or classifier outputs, and nothing commits that customer images will not be used to train or refine IQ UIP or future models. The privacy policy allows some transfers across jurisdictions.
Business associate status is stated in a posted agreement for the Imbio portal. Imbio acts as a business associate when it processes protected health information for covered entities, and Exhibit A of its portal agreement binds subcontractors to the same restrictions, requires reporting certain incidents within 72 hours and provides for return or destruction of the information at termination where feasible.
That agreement is written for the Imbio portal products. 4DMedical's own commercial terms put de identifying images before transmission and re identifying results locally on the customer, and no business associate agreement is posted for CT:VQ, CT LVAS or deployments through the 4DMedical platform, so the coverage across the suite is not stated.
Security commitments are contractual. Imbio's business associate agreement requires safeguards for protected health information, passes restrictions to subcontractors and requires reporting certain incidents within 72 hours, the CT LVAS workflow places a DICOM router under the institution's own security, a contact exists for reporting vulnerabilities, and access to personal information is limited to staff with a business need.
No SOC 2 report, ISO 27001 certificate, HITRUST assessment or penetration test is claimed for 4DMedical or its hosted service, and the device quality system covers development and risk controls rather than operations. How the cloud service handling chest CTs is run has not been examined by an outside auditor.
Both AI relevant clearances are identifiable. FDA cleared IQ UIP under K242467 in December 2024 to notify specialists at interstitial lung disease centers of adult noncontrast chest CT findings suggestive of usual interstitial pneumonia, as an informational worklist notice that does not confirm a diagnosis, and cleared CT:VQ under K251484 in August 2025 as decision support for adult thoracic disease using regional ventilation and perfusion from CT; CT:VQ also holds clearances in the EU, UK, Canada, Australia and New Zealand.
The site's menu labels IQ UIP as early IPF detection, wording broader than the cleared notification of a suspicious pattern, and lung texture analysis is marked investigational only in the United States.
IQ UIP's July 2026 instructions for use compare its performance across groups. On 804 exams kept apart from training and tuning, sensitivity and specificity are reported by lung disease, scanner manufacturer, reconstruction kernel and slice thickness; at the operating point, sensitivity was 82.2% on GE images and 95.1% on Philips images, with specificity of 94.1% and 77.4% respectively, and the instructions warn users about the lower specificity on Philips scans.
That lets a department see how its own scanners will affect alerts. Results by age, sex or race are not given, and CT:VQ and the other tools, which claim no learned model, carry no comparable breakdown.
IQ UIP's error rate is published with its method and denominator. The K242467 validation tested 804 chest CT exams against a thoracic radiologist reference, and at the 0.395 threshold sensitivity was 90.2% with a 95% confidence interval of 86.2% to 94.3% and specificity 91.5% with an interval of 89.2% to 93.7%, so a buyer knows roughly how many alerts will be false and how many cases will be missed. The reviewing team keeps the original scan and can disagree.
No remedy attaches to a wrong alert. The Imbio portal agreement limits liability except for Imbio's own infringement of intellectual property, and no correction route runs to a patient whose scan was missed.
Integration runs through imaging standards with named partners. IQ UIP takes high resolution noncontrast chest CT in DICOM and sends positive findings to a separate specialist worklist while the source exam stays in the clinical PACS, CT LVAS sends scans and returns reports through a DICOM router the institution manages, and CT:VQ returns quantitative images and DICOM structured reports to PACS. Laurel Bridge Software and Kailo Medical package the data exchange, and CT:VQ runs at UC San Diego Health, Cleveland Clinic and Stanford.
No PACS, RIS or EHR product is named with its supported read and write operations, and results reach the chart only through the radiology report.
Processing is hosted. Every product is listed as software as a service, scans leave the provider through a locally managed DICOM router or a packaged connection, and IQ UIP can also be deployed through distribution partners or the 4DMedical Protected Platform, with notifications reaching clinicians on the hospital network.
No region or country is committed for image storage or model processing, and the privacy policy allows some transfers across jurisdictions, which matters for a company headquartered in Australia with US operations. Customers de identify images before transmission, which limits what leaves the site without fixing where it is processed.
The products are sold as software as a service on contract. The XV Technology business charges a fee per scan, mostly from CT LVAS, according to 4DMedical's 2025 investor presentation, and CT:VQ is introduced through a clinical demonstration and sample report, with availability varying by country.
No price per scan, subscription, minimum or volume tier is published for CT:VQ, CT LVAS, IQ UIP or the other tools. The Category III CPT codes that let US providers bill eligible scans are reimbursement to the provider, not what the provider pays 4DMedical, so a department can see that cost follows scan volume but cannot size it before contact.
Each product names its setting. IQ UIP analyzes qualifying adult noncontrast chest CTs from a hospital network, including standalone imaging clinics tied to it, and notifies clinicians at an interstitial lung disease center, within acquisition limits on kernel, slice thickness and series. CT:VQ serves radiologists, pulmonologists and nuclear medicine physicians assessing adult thoracic disease, as a noncontrast alternative to nuclear V/Q.
Validation backs both: IQ UIP's 804 exam cohort spans several lung diseases and scanner makers, and CT:VQ's submission compared it with SPECT, with named academic sites in use. Lung texture analysis is for investigational use only in the United States, and the newer calcification and pulmonary hypertension tools show no validation on the product pages.
Alternatives to 4DMedical
Other products that do the same kind of work as 4DMedical, and what each one does.
Oxford spinout that sells AI imaging software for stroke and lung disease.
contextflow reads chest CT and shows the radiologist comparable cases alongside its findings.
CureMetrix, a San Diego company founded in 2014, sells AI software that reads screening mammograms for breast imaging centers, mammography practices and hospital radiology departments.
Beijing based imaging AI company whose InferRead line detects findings on CT and radiographs (lung nodules on chest CT, chest abnormalities on radiographs, coronary stenosis on CTA, hemorrhage on head CT, chest fractures and pneumonia), alongside the InferOperate suite of 3D…
ANDI, short for Advanced Neuro Diagnostic Imaging, is brain MRI analysis software from Imeka Solutions in Sherbrooke, Quebec.
Caristo Diagnostics, an Oxford spinout, measures coronary inflammation from the same routine coronary CT angiogram used to assess plaque.
Pricing
Vendor-published figures are labeled as such. Figures labeled “Estimated” are derived from third-party sources and have not been confirmed by the vendor.
| Entry Price | Pricing Basis | BAA Tier | Implementation | Source |
|---|---|---|---|---|
|
Commercial quote
|
Software as a service on contract. XV Technology charges a fee per scan, mostly from CT LVAS, and CT:VQ, CT LVAS and IQ UIP configurations are quoted by the imaging sales team. No rates are published. | Imbio offers a business associate agreement for covered entity customers using its portal. The broader pulmonary imaging contract determines how protected information is handled for other products. | Deployment can require a locally managed DICOM router and setup for report delivery into the imaging workflow. Installation and support fees depend on the configuration quoted. | Vendor Published |
Eligible US scans can be billed by the provider under Category III CPT codes; that reimbursement is separate from the software fee. Lung texture analysis is not for commercial sale in the United States.