Velacur ONE
Velacur ONE is a portable ultrasound elastography system from Sonic Incytes Medical Corp., founded in Vancouver in 2017. It measures liver stiffness, ultrasound attenuation and Velacur Determined Fat Fraction. An activation pad sends shear waves through the tissue while the operator moves a single handheld probe across the liver. The system calculates stiffness from the shear waves and estimates fat fraction from attenuation and backscatter. The Organ Guide adds a color overlay to help locate the liver. The Wave Quality Guide identifies usable waves, and an on screen warning flags scans with low image or tracking quality. Trained clinicians interpret the measurements with the patient's other findings.
The FDA cleared Velacur ONE model LI-1100 in June 2025 under K251728. Its interface and hardware changed, while the indication and measurement workflow carried forward from the original Velacur. The Norfolk, Virginia, location of Capital Digestive Care reported more than 2,500 procedures in about a year after installing the earlier device, with procedure volume rising 171%. Its result predates Velacur ONE. Sonic Incytes announced a multiyear partnership with Madrigal Pharmaceuticals in November 2026 to expand point of care liver assessment in gastroenterology, hepatology and endocrinology. Clinics buy the device through Sonic Incytes, which also provides clinical training and technical support.
Capability Axes
An AI Health Index grade measures what a buyer can verify from public sources on the date shown. It is not a rating of how good the product is. A vendor can build an excellent system and grade low on an axis because it publishes nothing an outsider can check. How grades read
The scan measures liver stiffness from mechanically generated shear waves and measures attenuation from the loss of power in the ultrasound signal. The system combines attenuation and backscatter in a fitted calculation for fat fraction. Its learned models help the operator acquire usable images. The Organ Guide marks the liver on the image and can be switched off, and the Wave Quality Guide identifies the parts of the field in which the waves are usable.
The probe and activation pad still acquire the wave field and return stiffness and attenuation without the model based guidance. Removing the learned components would also remove the fitted fat fraction estimate. The underlying ultrasound elastography remains a functioning liver assessment instrument.
The operator places the activation pad under the patient's back and sweeps the probe through the field, using the B mode image to locate the liver. The optional Organ Guide overlays a color boundary without taking over acquisition. The Wave Quality Guide marks areas with usable waves, and the display shows an image or tracking quality warning when acquisition quality falls too low. A trained person controls the probe and decides whether to repeat a scan.
The device calculates the measurements and the clinician interprets them with other clinical indicators. The user interface going into the ONE release received formative and summative evaluation. No escalation or correction workflow covers a misleading organ boundary or a clinically implausible result.
The system uses shear wave speed to calculate tissue stiffness and combines attenuation and backscatter for its fat fraction estimate. Its optional Organ Guide and Wave Quality Guide use machine learning to identify anatomy and usable wave patterns. The predicate model filing describes training the organ guide on more than 5,000 images and evaluating it on more than 800 images from 21 patients.
It describes a separate wave guide training set of more than 15,000 images and an evaluation of more than 4,000 images from 36 patients. The ONE release preserves the guidance functions. Buyers get no build identifier for either guide and no conditions for replacing or updating them during the device's service life.
Sonic Incytes packages the ultrasound acquisition, computing and the guidance tools in the Velacur ONE system. The Organ Guide analyzes anatomy on the device and the Wave Quality Guide identifies usable regions as the scan takes place. The learned methods use ultrasound images from volunteers and patients and were assessed against expert segmentation in separate evaluation groups during development of the predicate.
In routine use the overlay and measurement appear on the local console, which also controls the activation source and processes the incoming ultrasound data. The detection path does not require an outside model service for that step. The service and support offer has no named external processor or data terms for anyone who might receive an exported scan for troubleshooting.
Capital Digestive Care's Norfolk site reported a 171% increase in procedure volume after adding the original Velacur and more than 2,500 studies over about a year. A peer reviewed 2025 study comparing the earlier unit with biopsy included 70 of 78 enrolled adults and reported an area under the curve of 0.846 for moderate steatosis using fat fraction. Velacur ONE retains the measurement outputs and introduces new hardware and a revised interface.
Its FDA submission uses bench comparisons on tissue mimicking phantoms and formative and summative usability testing for the new hardware. There is no named clinical deployment of the ONE unit with a measured outcome comparable to the earlier Norfolk case study.
The Organ Guide and Wave Quality Guide operate while a clinician acquires the ultrasound. The device shows a quality warning and retains the normal image view, so the operator can see the tissue and determine whether to continue acquisition. Sonic Incytes uses separate training and evaluation sets in developing the learned guides, and the earlier filing describes expert segmentation and separation of validation data by patient and site.
Patient information on the clinical console has no deletion deadline and no restriction on model training, the only privacy terms cover website visitors, and no incident process covers a recorded exam or a misleading measurement.
The patient exam runs on an ultrasound console in a clinic or hospital. The operator sees the image and measurements on the instrument, and the clinician interprets them with the rest of the patient's findings. Reimbursement documentation for an imaging exam includes the patient's identity, facility, date and results in the provider's medical record. Sonic Incytes collects contact information for demonstrations and provides separate training and technical support.
The company privacy terms govern information visitors submit through its website. They do not cover patient scans processed by the console or the handling of a case during support. The equipment offer contains no business associate status or equivalent processor terms for those images.
The redesigned ultrasound system consolidates computing and control functions and runs the measurement workflow on the device. For the ONE release, the FDA filing identifies device testing against IEC 81001 5 1, NIST SP 800 53 and standards and guidance for medical device network security. The testing included functional, boundary and stress checks. The review assessed the unit alongside comparison with the predicate and testing of the new hardware and software design. Beyond that premarket device review, no SOC 2 report or information security certificate covers Sonic Incytes's own handling of scans or its support services.
The FDA cleared Velacur ONE model LI-1100 under K251728 on June 26, 2025. It estimates liver tissue stiffness, ultrasound attenuation and Velacur Determined Fat Fraction for use by trained professionals in a clinical setting. The fat fraction estimate is not intended for pediatric patients. The readings are for use alongside other clinical indicators in the management of liver disease, including hepatic steatosis.
The ONE model updates the computing unit, activation pad and interface while preserving the predicate Velacur device's indication and measurement workflow. The predicate's 2024 K233977 added the fat fraction output. The marketed references to MASLD and MASH concern use of those quantitative measurements for clinical assessment.
The optional Organ Guide marks liver anatomy and the Wave Quality Guide highlights areas in which the scan shows usable waves. The earlier model filing included volunteers and people with more severe liver disease from hepatology clinics, with evaluation datasets kept separate from training and checked against expert drawn regions.
Sonic Incytes sells the AI guidance as enabling confident measurement even with inexperienced operators or high BMI patients, and Velacur ONE adds depth aimed at that group. No results are reported by BMI group, operator experience or disease stage for the guides or for usable scan rates on the ONE release, and the prior Capital Digestive Care case, with patients from BMI 16.3 to 50.1, pools its results across that range.
The trained operator keeps control of the scanner, can switch off the Organ Guide and sees the underlying ultrasound with the model output overlaid. A quality warning gives the operator another chance to repeat an inadequate acquisition. The measurements supplement the clinician's other evidence, and the individual report supports review of what the scanner returned. Sonic Incytes provides a technical support channel and separate clinical training for the device.
Training resources cover the placement of the pad and the sequence used to obtain an exam. Its online terms govern use of the company website. The equipment offer has no specific obligation to correct a wrong liver boundary or an inaccurate fat fraction measurement, and the support channel does not establish compensation or an output review remedy.
The operator uses the device's local display for the ultrasound view and guidance overlays. The clinician sees quantitative stiffness and attenuation along with the fat fraction measure once the scan is complete, so the patient can be counseled in the same visit. The reimbursement workflow calls for a written report and images to be kept as part of the patient record, including the date and findings.
The console gives the clinician the measurements by the end of the visit, and Sonic Incytes provides training for equipment operators. The equipment offer has no named interface that writes the measurements or images to a radiology archive or an EHR, and no receiving system is named for the report.
The system runs in the exam room on a compact console linked to the activation pad and handheld probe. The ONE design combines computing and control functions and displays guidance and results locally while the clinician is with the patient. The probe and activation source stay at the care site, and image acquisition and measurement occur in that installation. The console's security was evaluated for network operation as part of the regulatory submission.
Sonic Incytes offers technical support and clinical training. The onsite system fixes the location of the ultrasound and inference, but does not specify where longitudinal exam copies are stored or whether technical support can access them outside the facility.
Sonic Incytes sells Velacur ONE as clinical equipment with an activation pad, ultrasound probe and integrated console, bought by the clinic after a demonstration. The purchase includes clinical training and technical assistance, and the equipment is designed for office use, producing stiffness, attenuation and fat fraction in the exam room.
Sonic Incytes markets a lower upfront cost and higher reimbursement than transient elastography without imaging, with no amount attached to either claim. Providers document the exam for CPT 76981, and at Capital Digestive Care payers reimbursed 90% of the procedures billed under that code in the first 15 months with the original Velacur. Those are payer figures. The device price, financing terms and service charge come from sales.
The portable unit is intended for trained medical professionals in clinics and hospitals, with no sonography experience required. It measures stiffness, attenuation and fat fraction as aids for clinical assessment of liver disease. Gastroenterology and hepatology are core settings, and the Madrigal partnership also includes endocrinology. The original Velacur was used at Capital Digestive Care in Norfolk and studied alongside biopsy in adults with metabolic liver disease.
The ONE probe and shear wave design reach up to twice the measurement depth of conventional transient elastography. The same product page puts the single probe's range of body types at BMI 25 to 45 in one place and BMI 25 to 43 in another. Velacur Determined Fat Fraction is intended for adults, and the ONE unit has no separate prospective assessment in a setting outside the original use.
Alternatives to Velacur ONE
Other products that do the same kind of work as Velacur ONE, and what each one does.
Vscan Air is GE HealthCare's handheld ultrasound: a wireless dual probe that pairs to a phone or tablet and fits in a coat pocket.
EchoNous makes Kosmos, a handheld ultrasound system positioned against cart based systems rather than pocket scanners.
Exo, pronounced echo, makes the Exo Iris handheld ultrasound probe and describes itself as a medical imaging software and devices company.
Philips Lumify is a handheld ultrasound transducer that plugs into a smart device the customer already owns, over USB on Android or Lightning on Apple hardware, with the app downloaded from the ordinary app stores.
MEDICAL IP builds AI segmentation and quantification software for medical imaging.
MediView XR builds XR90, an augmented reality visualization and navigation system for image guided interventional procedures, and markets OmnifyXR for interventional suites.
Pricing
Vendor-published figures are labeled as such. Figures labeled “Estimated” are derived from third-party sources and have not been confirmed by the vendor.
| Entry Price | Pricing Basis | BAA Tier | Implementation | Source |
|---|---|---|---|---|
|
Contact Sonic Incytes
|
Velacur ONE is purchased through Sonic Incytes after a demonstration of the ultrasound system and guidance tools. Device pricing and any finance or service arrangement are part of the facility discussion. | The console handles ultrasound exams in a clinical setting. The purchase materials do not establish a business associate agreement for technical support or other access to patient scans. | Sonic Incytes offers clinical training, equipment setup resources and technical support. A separate setup or training fee is part of the facility purchase terms. | Vendor Published |
The console estimates liver stiffness, attenuation and fat fraction. The CPT 76981 amounts in the billing guide are provider reimbursements for an exam. Equipment and service fees are handled in the facility purchase discussion.